FDA Recall Terminated

Asahi APS Series Dialyzers, Model Nos. APS-400S, APS-550S, APS-650S, APS-750S, APS-900S, APS-1050S; the dialyzers are intended for single or initial use and under reprocessing and reuse conditions fo rup to 15 reuse cycles; Manufacturer: Asahi Medical Co., Ltd., 9-1, Kanda Mitoshirocho, Chiyoda-ku, Tokyo 101-8482, Japan; Distributed in teh USA by Nextron Medical, 45 Kulick Road, Fairfield, NJ 07004 USA

Recall: Z-0739-04 · Initiated October 1, 2001

Recall

Recall Number
Z-0739-04
Event Number
28341
Firm
Asahi Medical Co Ltd 1-1 Uchiauwaicho, 1-Chome 14 Flr, Imperial Tower Bldg Chiyoda Ku Japan
FEI Number
3002806407
Product Code
KDI
Status
Terminated
Root Cause
Other
Initiated
October 1, 2001
Posted
July 20, 2004
Terminated
March 24, 2004

Description

Asahi APS Series Dialyzers, Model Nos. APS-400S, APS-550S, APS-650S, APS-750S, APS-900S, APS-1050S; the dialyzers are intended for single or initial use and under reprocessing and reuse conditions fo rup to 15 reuse cycles; Manufacturer: Asahi Medical Co., Ltd., 9-1, Kanda Mitoshirocho, Chiyoda-ku, Tokyo 101-8482, Japan; Distributed in teh USA by Nextron Medical, 45 Kulick Road, Fairfield, NJ 07004 USA

Reason

Customer reprocessing methods for the reuseable hollow fiber dialyzers may damage the hollow fibers and result in blood leaks during dialysis.

Action

Asahi visited each of the U.S. customers starting on 10/1/01, observing how the facility handles the dialyzer before pre-cleaning, during pre-cleaning and priming. The users were retrained in the reprocessing steps, emphasizing the controls and care to be taken during pre-cleaning, including the verification of water pressure.

Distribution

New Jersey, Ohio, Texas, California, Georgia, Pennsylvania, Illinois and Indiana

Quantity

180,000 units estimated