Carbomedics VT-200 Valve Tester
Recall
- Recall Number
- Z-0738-04
- Event Number
- 28335
- Firm
- Carbomedics, Inc
- FEI Number
- 1627803
- Product Code
- DTJ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 30, 2002
- Posted
- July 20, 2004
- Terminated
- March 24, 2004
- Address
- 1300 E Anderson Ln, Austin, TX, 78752-1708
Description
Carbomedics VT-200 Valve Tester
Excessive cleaning and resterilization cause cracking and crazing of device causing the device to malfunction during use.
Letters were sent to surgeons and hospital operating room supervisors in mid-May 2002 which included the following information: It is imperative that hospitals regularly and thoroughly inspect all testing instruments for indications of wear and degredation. And, CarboMedics is in the process of re-issuing suggested maintenance and sterilization guidlines.
US nationwide and to the following foreign countries: Denmark, Ireland, China, Cyprus, Greece, France, UK, Protugal, Ecuador, India, Japan, Hong Kong, Italy, Sweden, Belgium, Norway, Argentina, Spain, Malta, Austria, Switzerland, Netherlands, Germany, Australia, Finland
4606 units