FDA Recall Terminated

Carbomedics VT-200 Valve Tester

Recall: Z-0738-04 · Initiated April 30, 2002

Recall

Recall Number
Z-0738-04
Event Number
28335
Firm
Carbomedics, Inc
FEI Number
1627803
Product Code
DTJ
Status
Terminated
Root Cause
Other
Initiated
April 30, 2002
Posted
July 20, 2004
Terminated
March 24, 2004
Address
1300 E Anderson Ln, Austin, TX, 78752-1708

Description

Carbomedics VT-200 Valve Tester

Reason

Excessive cleaning and resterilization cause cracking and crazing of device causing the device to malfunction during use.

Action

Letters were sent to surgeons and hospital operating room supervisors in mid-May 2002 which included the following information: It is imperative that hospitals regularly and thoroughly inspect all testing instruments for indications of wear and degredation. And, CarboMedics is in the process of re-issuing suggested maintenance and sterilization guidlines.

Distribution

US nationwide and to the following foreign countries: Denmark, Ireland, China, Cyprus, Greece, France, UK, Protugal, Ecuador, India, Japan, Hong Kong, Italy, Sweden, Belgium, Norway, Argentina, Spain, Malta, Austria, Switzerland, Netherlands, Germany, Australia, Finland

Quantity

4606 units