FDA Recall
Terminated
ACL Futura Instrument Analyzer
Recall: Z-0736-04
·
Initiated March 4, 2004
Recall
- Recall Number
- Z-0736-04
- Event Number
- 28485
- Firm
- Instrumentation Laboratory Co.
- FEI Number
- 1217183
- Product Code
- JPA
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 4, 2004
- Posted
- July 20, 2004
- Terminated
- January 28, 2016
- Address
- 113 Hartwell Ave, Lexington, MA, 02421-3125
Description
ACL Futura Instrument Analyzer
Reason
Software may cause instrument to omit step causing reagent carrover which may effect patient test result
Action
Instrumentation Laboratory (IL) notified the Distributor Beckman Coulter, FL by letter on 3/9/04. Beckman will notify customers of the mandatory software upgrade by letter dated 3/4/04 with instructions for the software installation.
Distribution
Florida Canada
Quantity
1450 units