VITROS Chemistry Products VersaTips, Catalog No. 6801715 A VersaTip sales unit is composed of 1000 tips per box.
Recall
- Recall Number
- Z-0735-2022
- Event Number
- 89362
- Firm
- Ortho-Clinical Diagnostics, Inc.
- FEI Number
- 1000136573
- Product Code
- JRC
- Status
- Open, Classified
- Root Cause
- Nonconforming Material/Component
- Initiated
- December 10, 2021
- Address
- 100 Indigo Creek Dr, Rochester, NY, 14626-5101
Description
VITROS Chemistry Products VersaTips, Catalog No. 6801715 A VersaTip sales unit is composed of 1000 tips per box.
The lot has a shape deformity which may lead to a 16% reduction in the volume of the sample or reagent dispensed. In cases where the error is not detected and the affected results are not suppressed, clinical impact may include reduced drug dosage, or false positive or false negative results for bloodborne infectious disease assays when sample results are close to the cut-off value.
On December 10, 2021, the firm notified consignees through ORTHO PLUSSM (a service that provides customers with Ortho Clinical Diagnostics product notifications electronically) of the product issue. Customers were instructed to discontinue use and discard the affected lot. Customers should forward the notification if the product was distributed outside of their facility.
Domestic distribution nationwide. Foreign distribution to Canada.
4125 units