FDA Recall Terminated

NucliSensEasy Q Incubator, Product Number 285204 (this number can be found at the back of the instrument).

Recall: Z-0735-04 · Initiated February 26, 2004

Recall

Recall Number
Z-0735-04
Event Number
28435
Firm
Biomerieux, Inc.
FEI Number
1039284
Product Code
JJQ
Status
Terminated
Root Cause
Other
Initiated
February 26, 2004
Posted
July 20, 2004
Terminated
March 24, 2004
Address
100 Rodolphe St, Durham, NC, 27712-9402

Description

NucliSensEasy Q Incubator, Product Number 285204 (this number can be found at the back of the instrument).

Reason

Instrument malfunction. Electrical short circuit may occur resulting in damage or a fire hazard.

Action

This correction is being handled as an instrument repair (rework) by bioMerieux Field Service; therefore no communication has been sent to the customer sites. The repair involves heat sink replacement or use of insulating tape on the rear side of the keypad.

Distribution

CT, MD, MT

Quantity

3 units