FDA Recall Terminated

Sechrist Hyperbaric Chamber Gurney, Standard Gurney & Deluxe Wide Stretcher Model, P/N 21464, Sechrist Industries, Inc., Anaheim, CA

Recall: Z-0734-2008 · Initiated October 31, 2007

Recall

Recall Number
Z-0734-2008
Event Number
45906
Firm
Sechrist Industries Inc
FEI Number
2020676
Product Code
FPO
Status
Terminated
Root Cause
Device Design
Initiated
October 31, 2007
Posted
January 18, 2008
Terminated
April 19, 2012
Address
4225 E La Palma Ave, Anaheim, CA, 92807-1815

Description

Sechrist Hyperbaric Chamber Gurney, Standard Gurney & Deluxe Wide Stretcher Model, P/N 21464, Sechrist Industries, Inc., Anaheim, CA

Reason

Brake Problems: The stretcher brake cams may crack while attempting to lock the stretcher brake (Sechrist is supplied the stretchers by the manufacturer Stryker Medical)

Action

Urgent Medical Device Correction letters were sent to customers beginning on November 9, 2007, via FedEx & DHL overnight. Customers were notified this action was being taken to providing notification to their customers as a continued implementation of the Stryker Medical field correction. The Notification letter instructed their customers to: 1. Locate the gurneys with the serial numbers listed in this advisory. 2. If any of the affected gurneys cannot be located, determine if they have been sold, transferred, or donated to another facility and notify us regarding their new location. 3. Verify the brakes are working properly. The brake function may be checked by engaging the brake pedal and pushing on each end of the gurney to make sure that casters are locked securely. 4. Remove the gurneys with non-functioning brakes from service until brake cams have been replaced and brakes are fully functional. 5. Contract Sechrist for the replacement parts package. 6. Complete the retrofit of the Gurney. 7. Verify correct function of the brake assembly following retrofit. 8. Complete and send the Correction Verification Form contained in the package to Sechrist.

Distribution

Worldwide-USA, Brazil, Bulgaria, Canada, Dubai, U.A.E., Estonia, India, Indonesia, Japan, Malaysia, Saudi Arabia, Sweden, Taiwan, Thailand, and Vietnam.

Quantity

559 Units Total (483 in US) for all models