FDA Recall Terminated

Vital Diagnostics Envoy Calcium Reagent For In-Vitro Diagnostic Use. PN 55285 For the quantitative measurement of calcium in plasma and serum

Recall: Z-0732-2014 · Initiated August 29, 2013

Recall

Recall Number
Z-0732-2014
Event Number
66202
Firm
Vital Diagnostics, Inc.
FEI Number
1717966
Product Code
CJY
Status
Terminated
Root Cause
Device Design
Initiated
August 29, 2013
Posted
January 14, 2014
Terminated
August 26, 2015
Address
27 Wellington Rd, Lincoln, RI, 02865-4411

Description

Vital Diagnostics Envoy Calcium Reagent For In-Vitro Diagnostic Use. PN 55285 For the quantitative measurement of calcium in plasma and serum

Reason

High positive calcium bias on plasma sample versus results on serum

Action

Vital Diagnostics issued a Recall Notice dated August 29, 2013, to all affected customers via certified mail. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed not to use plasma for the determination of total calcium when using the affected product. The use of serum is acceptable, thus the product can still be used to determine total calcium with this sample type. Vital Diagnostics released a corection to the package insert for the Envoy Calcium Reagent. Customers with the affected product in their laboratory, were instructed to discard the insert packed with the product and replace it with the insert enclosed with this bulletin. If product was transferred to another location, the bulletin should be forwarded to that location. The package insert was revised indicating to only use Serum samples when running Calcium. Vital Diagnostics issued a revised (expanded) notice dated September 19, 2013, to include the Consult Diagnostics Eon Calcium Reagent (PN: 1429) requesting users to discontinue using the single vial Eon Calcium kit (PN: 1429). Customers with questions were instructed to call Vital Diagnostics Technical Support at 1-855-354-8324.. For questions regarding this recall call 401-642-8400.

Distribution

Nationwide Distribution

Quantity

395 units