FDA Recall Open, Classified

(1) Patient Information Center iX, Model No. 866389 (2) Patient Information Center iX Expand, Model No. 866390 (3) PIC iX Hardware, Model No. 866424

Recall: Z-0731-2024 · Initiated November 28, 2023

Recall

Recall Number
Z-0731-2024
Event Number
93638
Firm
Philips North America
FEI Number
3006648320
Product Code
MHX
Status
Open, Classified
Root Cause
Software design
Initiated
November 28, 2023
Posted
January 13, 2024
Address
222 Jacobs St, Cambridge, MA, 02141-2296

Description

(1) Patient Information Center iX, Model No. 866389 (2) Patient Information Center iX Expand, Model No. 866390 (3) PIC iX Hardware, Model No. 866424

Reason

Push notifications may fail to send to the user under certain conditions. This could potentially result in patient harm due to delay in detection of a change in patient condition.

Action

On November 28, 2023, the firm notified affected consignees through URGENT Medical Device Correction letters. This failure is applicable to Apple iOS devices only. If customers use iOS devices (e.g. iPhone, iPad, etc.) to receive event notifications: 1) Please conduct patient monitoring at the central station or patient bedside. 2) Please do not use iOS event notifications for patient monitoring. A Philips representative will contact you to schedule a visit to install a software patch to your device which will correct the current issue. Patch PIC iX 4.2.2 will be provided for PIC iX 4.X and patch CareEvent C.03.07 will be provided for CareEvent C.03.X. If you need any further information, please contact your local Philips representative, or call 1-800-722- 9377.

Distribution

Worldwide distribution: US (Nationwide) to states of: CA, GA, IA, MA, MI, MN, NJ, NM, NY, OH, PA, TN, TX, WA, WV; and OUS (Foreign) distribution to countries of: Australia, Austria, Belarus, Belgium, Brazil, Canada, Czech Republic, Denmark, Estonia, Faroe Islands, Finland, France, Germany, India, Indonesia, Ireland, Italy, Japan, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Poland, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Kingdom, Utd.Arab.Emir., and Uzbekistan

Quantity

41 US; 773 OUS