PKG, ASPIRATION NEEDLE, 45 CM, P/N 0250080703 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Recall
- Recall Number
- Z-0731-2015
- Event Number
- 69838
- Firm
- Stryker Endoscopy
- FEI Number
- 2936485
- Product Code
- KOG
- Status
- Terminated
- Root Cause
- Error in labeling
- Initiated
- November 17, 2014
- Posted
- December 12, 2014
- Terminated
- July 17, 2015
- Address
- 5900 Optical Ct, San Jose, CA, 95138-1400
Description
PKG, ASPIRATION NEEDLE, 45 CM, P/N 0250080703 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to [email protected] or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at [email protected]. For questions regarding this recall call 408-754-2000.
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
180,573 units total