FDA Recall Open, Classified

B105P Patient Monitor (with E-module slot option), REF 6160000-001-XXXXXX,

Recall: Z-0729-2023 · Initiated November 16, 2022

Recall

Recall Number
Z-0729-2023
Event Number
91176
Firm
GE Medical Systems, LLC
FEI Number
2126677
Product Code
MHX
Status
Open, Classified
Root Cause
Device Design
Initiated
November 16, 2022
Posted
December 19, 2022
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

B105P Patient Monitor (with E-module slot option), REF 6160000-001-XXXXXX,

Reason

GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that is not at or near sea level.

Action

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/16/2022. The notice explained the problem and safety issue and requested the following: "You can determine whether your monitor is impacted by looking at the side of your monitor where the parameter acquisition connectors are located. B1x5P with E-module slot option and all B1x5M patient monitors are impacted (See Figure 1). B1x5P patient monitors without E-module slot option are not impacted (See Figure 2). You can also determine whether your monitor is impacted following the affected product directions below. You can continue to use the affected monitors by following the safety instructions below. 1. Use CO2 measurement in % unit by following the below steps on the B1x5P / B1x5M monitor. a. From the Main Screen, click the CO2 digit field; b. Select Setup > Unit; c. Select % from the dropdown list. OR If available, you can also insert the GE Healthcare gas E-module for CO2 measurement directly into the E-module slot of the GE Healthcare anesthesia delivery systems (Carestation 600 Series, Carestation 750, Avance CS2, and Aisys CS2) or CARESCAPE R860 ventilator, as the CO2 measurement values displayed on these machines are correct in all "%", mmHg , and kPa units." "If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative." The affected products will be updated.

Distribution

Worldwide distribution.

Quantity

826 devices