FDA Recall Terminated

Invictus-- Bionime Rightest Blood Glucose Monitoring System MODEL: GM300; Invictus Scientific, Inc., San Diego, CA

Recall: Z-0728-2008 · Initiated November 26, 2007

Recall

Recall Number
Z-0728-2008
Event Number
45953
Firm
Invictus Scientific Inc
FEI Number
3005498483
Product Code
NBW
Status
Terminated
Root Cause
Software design
Initiated
November 26, 2007
Posted
February 16, 2008
Terminated
October 13, 2011
Address
11760 Sorrento Valley Rd, Ste G, San Diego, CA, 92121

Description

Invictus-- Bionime Rightest Blood Glucose Monitoring System MODEL: GM300; Invictus Scientific, Inc., San Diego, CA

Reason

Abnormal Memory Recall: When the meter is fixed in 12 hour mode, the time frame display of all memory data will be shown incorrectly as AM or PM depending on when the user is recalling the memory stored in the meter. This will cause confusion as to when the last test was taken.

Action

An Urgent: Device Recall Notification letter were sent to customers (direct accounts) beginning on November 20, 2007, via US Postal Service, with a return receipt requested. The distributors were instructed to inform consumers/users to fix the time format to the 24 hrs mode and to contact the distributors or the firm for replacement, if they prefer to use the time format in the 12 hrs mode. They were also instructed that if they have further distributed any of the recalled products, to please immediately contact those accounts and advise them of the recall situation, and have them return their outstanding recalled stocks. If customers have any questions, they were instructed to contact at 858-481-8485.

Distribution

Nationwide Distribution

Quantity

21,070