ProxiDiagnost N90 Image-intensified fluoroscopic x-ray system
Recall: Z-0727-2021
·
Initiated December 8, 2020
Recall
- Recall Number
- Z-0727-2021
- Event Number
- 87039
- Firm
- PHILIPS HOME HEALTHCARE SOLUTION
- FEI Number
- 3016860078
- Product Code
- OWB
- Status
- Terminated
- Root Cause
- Radiation Control for Health and Safety Act
- Initiated
- December 8, 2020
- Posted
- January 28, 2021
- Terminated
- June 21, 2024
- Address
- 222 Jacobs St, Cambridge, MA, 02141-2289
Description
ProxiDiagnost N90 Image-intensified fluoroscopic x-ray system
Reason
After the user selects the Lock-in command, the kV and mA values are not locked immediately and can adjust if the collimator shutter position is changed directly after the Lock-in command.
Distribution
Nationwide
Quantity
91