PatientNet Monitoring System - PatientNet Central Stations utilizing software Versions 1.03.14, 1.04xx, 2.00xx.

Recall: Z-0727-04 · Initiated February 27, 2004

Recall

Recall Number
Z-0727-04
Event Number
28410
Firm
General Electric Medical Systems Information Technology
FEI Number
2124823
Product Code
MSX
Status
Terminated
Root Cause
Other
Initiated
February 27, 2004
Posted
July 20, 2004
Terminated
November 17, 2005
Address
8200 W Tower Ave, Milwaukee, WI, 53223-3219

Description

PatientNet Monitoring System - PatientNet Central Stations utilizing software Versions 1.03.14, 1.04xx, 2.00xx.

Reason

Non-invasive blood pressure readings on central station may not match bedside monitor due to software anomaly.

Action

Consignees were notified via certified mail, return receipt requested, on Friday, February 27, 2004. GE Field service representatives will upgrade affected PatientNet Central Stations (RVS, IRVS, or Sitelink IRVS) to software version 2.01.

Distribution

Nationwide throughout the United States

Quantity

Undetermined a this time. 574 sofware units were distributed with ability to upgrade one or multiple devices.