PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MM REGULAR/1975.14.500, BASEPLATE D.28MM REGULAR/1975.14.800, BASEPLATE D.25MM FULL WEDGED 10-degrees/1975.14.510, BASEPLATE D.28MM FULL WEDGED 10-degrees/1975.14.810, BASEPLATE D.25MM FULL WEDGED 15-degrees/1975.14.615, BASEPLATE D.28MM FULL WEDGED 15-degrees/1975.14.815, BASEPLATE D.28MM FULL WEDGED 15-degrees X/1975.14.865, BASEPLATE D.25 MM FULL WEDGED 20-degrees/1975.14.620, BASEPLATE D.28 MM FULL WEDGED 20-degrees/1975.14.820, BASEPLATE D.28 MM FULL WEDGED 20-degrees X/1975.14.870

Recall: Z-0726-2024 · Initiated August 29, 2023

Recall

Recall Number
Z-0726-2024
Event Number
93668
Firm
Limacorporate S.p.A Via Nazionale 52 Di San Daniele San Daniele Del Friuli Italy
FEI Number
3008021110
Product Code
MBF
Status
Open, Classified
Root Cause
Device Design
Initiated
August 29, 2023
Posted
January 12, 2024

Description

PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MM REGULAR/1975.14.500, BASEPLATE D.28MM REGULAR/1975.14.800, BASEPLATE D.25MM FULL WEDGED 10-degrees/1975.14.510, BASEPLATE D.28MM FULL WEDGED 10-degrees/1975.14.810, BASEPLATE D.25MM FULL WEDGED 15-degrees/1975.14.615, BASEPLATE D.28MM FULL WEDGED 15-degrees/1975.14.815, BASEPLATE D.28MM FULL WEDGED 15-degrees X/1975.14.865, BASEPLATE D.25 MM FULL WEDGED 20-degrees/1975.14.620, BASEPLATE D.28 MM FULL WEDGED 20-degrees/1975.14.820, BASEPLATE D.28 MM FULL WEDGED 20-degrees X/1975.14.870

Reason

Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue, which could lead to loosening of the implant, or the screw moving from its original position.

Action

On 8/29/23, the firm phoned all agents and surgeons to immediately stop surgeries with affected devices. On 10/10/23 the firm emailed recall notices to customers asking them to take the following actions: 1) Surgeons are suggested to perform a short-term patient s monitoring (radiographs at 6 weeks, 3-6 months and 1 year might be considered) and acting accordingly, based on a case-by-case evaluation. 2) Locate, quarantine, and return affected devices. Complete and return the response form to [email protected] 3) Customers with questions can email the firm at [email protected]

Distribution

US: IL, MO, CA, IN, MI, OK

Quantity

179