PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MM REGULAR/1975.14.500, BASEPLATE D.28MM REGULAR/1975.14.800, BASEPLATE D.25MM FULL WEDGED 10-degrees/1975.14.510, BASEPLATE D.28MM FULL WEDGED 10-degrees/1975.14.810, BASEPLATE D.25MM FULL WEDGED 15-degrees/1975.14.615, BASEPLATE D.28MM FULL WEDGED 15-degrees/1975.14.815, BASEPLATE D.28MM FULL WEDGED 15-degrees X/1975.14.865, BASEPLATE D.25 MM FULL WEDGED 20-degrees/1975.14.620, BASEPLATE D.28 MM FULL WEDGED 20-degrees/1975.14.820, BASEPLATE D.28 MM FULL WEDGED 20-degrees X/1975.14.870
Recall
- Recall Number
- Z-0726-2024
- Event Number
- 93668
- Firm
- Limacorporate S.p.A Via Nazionale 52 Di San Daniele San Daniele Del Friuli Italy
- FEI Number
- 3008021110
- Product Code
- MBF
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- August 29, 2023
- Posted
- January 12, 2024
Description
PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MM REGULAR/1975.14.500, BASEPLATE D.28MM REGULAR/1975.14.800, BASEPLATE D.25MM FULL WEDGED 10-degrees/1975.14.510, BASEPLATE D.28MM FULL WEDGED 10-degrees/1975.14.810, BASEPLATE D.25MM FULL WEDGED 15-degrees/1975.14.615, BASEPLATE D.28MM FULL WEDGED 15-degrees/1975.14.815, BASEPLATE D.28MM FULL WEDGED 15-degrees X/1975.14.865, BASEPLATE D.25 MM FULL WEDGED 20-degrees/1975.14.620, BASEPLATE D.28 MM FULL WEDGED 20-degrees/1975.14.820, BASEPLATE D.28 MM FULL WEDGED 20-degrees X/1975.14.870
Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue, which could lead to loosening of the implant, or the screw moving from its original position.
On 8/29/23, the firm phoned all agents and surgeons to immediately stop surgeries with affected devices. On 10/10/23 the firm emailed recall notices to customers asking them to take the following actions: 1) Surgeons are suggested to perform a short-term patient s monitoring (radiographs at 6 weeks, 3-6 months and 1 year might be considered) and acting accordingly, based on a case-by-case evaluation. 2) Locate, quarantine, and return affected devices. Complete and return the response form to [email protected] 3) Customers with questions can email the firm at [email protected]
US: IL, MO, CA, IN, MI, OK
179