CIC Pro Clinical Information Center Central Station - Product Usage: The intended use is to provide clinicians with adult, pediatric and neonatal patient data in a centralized location within a hospital or clinical environment. CIC Pro Clinical Information Center central station is intended to collect information from a network and display this data. This data includes physiological, patient demographic and/or other nonmedical information.
Recall
- Recall Number
- Z-0726-2020
- Event Number
- 84372
- Firm
- GE Healthcare, LLC
- FEI Number
- 2126677
- Product Code
- DSI
- Status
- Terminated
- Root Cause
- Software in the Use Environment
- Initiated
- November 12, 2019
- Terminated
- March 2, 2026
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
CIC Pro Clinical Information Center Central Station - Product Usage: The intended use is to provide clinicians with adult, pediatric and neonatal patient data in a centralized location within a hospital or clinical environment. CIC Pro Clinical Information Center central station is intended to collect information from a network and display this data. This data includes physiological, patient demographic and/or other nonmedical information.
When connected to the Mission Critical (MC) and /or Information Exchange (IX) networks, certain versions of the CARESCAPE Telemetry Server, ApexPro Telemetry Server, CARESCAPE Central Station (CSCS) version 1 and Central Information Center (CIC) systems were identified to have vulnerabilities to a cyber-attack.
The firm disseminated the notices by mail on 11/12/2019. The letter identified the security vulnerability of certain GE Central Stations and ApexPro Telemetry servers and provided additional best practices, ensuring the following: 1. MC and IX Networks are isolated; 2. MC and IX Router/Firewalls block incoming traffic, as applicable; 3. Restricted physical access to Central Stations, Telemetry Servers, MC network and IX network; 4. Default passwords are changed as applicable; and 5. Password management best practices are adhered to Ensuring the networks are properly configured and isolated protects against these potential concerns and mitigates the risk.
Worldwide distribution.
41848 units