FDA Recall Terminated

Magnesium on RX Imola analyser IVD

Recall: Z-0725-2017 · Initiated April 14, 2016

Recall

Recall Number
Z-0725-2017
Event Number
75263
Firm
Randox Laboratories, Limited Ardmore 55 The Diamond Road Crumlin United Kingdom
FEI Number
1000361607
Product Code
JGJ
Status
Terminated
Root Cause
Device Design
Initiated
April 14, 2016
Terminated
September 19, 2017

Description

Magnesium on RX Imola analyser IVD

Reason

According to the firm, Carry over was observed when the amylase or pancreatic amylase test is run directly before or after the Magnesium assay. An update is required to the RX Imola analyser running order and an acid wash recommended when setting up the magnesium assay. A correction was made to the IFU for the Magnesium assay and all RX Imola customers in the USA were contacted with the updated instruction.

Action

Randox sent an Urgent Field Safety Notice to all affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to review and update Rx Imola running order where required. Rework any remaining stock with updated IFU and Important Notice. Complete and return the vigilance response section of the form to [email protected].

Distribution

Nationwide Distribution The product was only distributed to US Consignees.

Quantity

15