FDA Recall Open, Classified

Medline brand, medical procedure convenience kits, labeled as: 1) NEURO CRANIOTOMY CDS #36-RF, SKU CDS780119T; 2) NEURO KIT, SKU CDS981327Q; 3) NEURO POSTERIOR SPINE PACK-LF, SKU DYNJ0311501I; 4) ACOUSTIC NEUROMA PACK-LF, SKU DYNJ37741D; 5) SCRIPPS SW NEURO PACK-LF, SKU DYNJ39836C; 6) SCRIPPS SW NEURO PACK-LF, SKU DYNJ39836D; 7) NEURO BUNDLE-LF, SKU DYNJ40095A; 8) NEURO BUNDLE-LF, SKU DYNJ40095B; 9) NEURO BUNDLE-LF, SKU DYNJ40095C; 10) NEURO BUNDLE-LF, SKU DYNJ40095D; 11) NEURO PACK-LF, SKU DYNJ42844D; 12) WMC VP SHUNT PACK-LF, SKU DYNJ50629J; 13) NEURO PACK, SKU DYNJ54255A; 14) NEURO CHRISTUS, SKU DYNJ61190A; 15) NEURO KIT, SKU DYNJ62854B; 16) NEURO KIT, SKU DYNJ62854C; 17) NEURO PACK NTX, SKU DYNJ66213D; 18) NEURO PACK, SKU DYNJ67628C; 19) BASIC NEURO CUSTOM, SKU DYNJ83387; 20) BASIC NEURO CUSTOM, SKU DYNJ83387A; 21) HALY NEURO PACK, SKU DYNJ87308; 22) FHWC NEURO, SKU DYNJ906131B; 23) CRANI, SKU DYNJ907865; 24) KIT NEURO LAFAYETTE, SKU DYNJ908615C; 25) GENERAL NEURO, SKU DYNJ909744

Recall: Z-0724-2025 · Initiated November 6, 2024

Recall

Recall Number
Z-0724-2025
Event Number
95806
Firm
MEDLINE INDUSTRIES, LP Northfield
FEI Number
1417592
Product Code
OJG
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
November 6, 2024
Posted
December 20, 2024
Address
3 Lakes Dr, Northfield, IL, 60093-2753

Description

Medline brand, medical procedure convenience kits, labeled as: 1) NEURO CRANIOTOMY CDS #36-RF, SKU CDS780119T; 2) NEURO KIT, SKU CDS981327Q; 3) NEURO POSTERIOR SPINE PACK-LF, SKU DYNJ0311501I; 4) ACOUSTIC NEUROMA PACK-LF, SKU DYNJ37741D; 5) SCRIPPS SW NEURO PACK-LF, SKU DYNJ39836C; 6) SCRIPPS SW NEURO PACK-LF, SKU DYNJ39836D; 7) NEURO BUNDLE-LF, SKU DYNJ40095A; 8) NEURO BUNDLE-LF, SKU DYNJ40095B; 9) NEURO BUNDLE-LF, SKU DYNJ40095C; 10) NEURO BUNDLE-LF, SKU DYNJ40095D; 11) NEURO PACK-LF, SKU DYNJ42844D; 12) WMC VP SHUNT PACK-LF, SKU DYNJ50629J; 13) NEURO PACK, SKU DYNJ54255A; 14) NEURO CHRISTUS, SKU DYNJ61190A; 15) NEURO KIT, SKU DYNJ62854B; 16) NEURO KIT, SKU DYNJ62854C; 17) NEURO PACK NTX, SKU DYNJ66213D; 18) NEURO PACK, SKU DYNJ67628C; 19) BASIC NEURO CUSTOM, SKU DYNJ83387; 20) BASIC NEURO CUSTOM, SKU DYNJ83387A; 21) HALY NEURO PACK, SKU DYNJ87308; 22) FHWC NEURO, SKU DYNJ906131B; 23) CRANI, SKU DYNJ907865; 24) KIT NEURO LAFAYETTE, SKU DYNJ908615C; 25) GENERAL NEURO, SKU DYNJ909744

Reason

Surgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those finished goods

Action

An IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification letter dated 11/7/24 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.comRecall Reference #: R-24-209-FGX1 Recall Code: Recall Code 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Distribution

US Nationwide. Canada.

Quantity

12,889 kits