Philips Allura Xper, UNIQ(Allura Xper system-China) Interventional Fluoroscopic X-ray System products with a Laird chiller WITHOUT an extended drip tray: Model Number: 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722029, 722033, 722034, 722036, 722038, 722039. Allura Xper F010 {C+F). 722003 Allura Xper F010/10 722005 Allura Xper F020 722006 Allura Xper F020/10 and F020/20 722008 Allura Xper F010 722010 Altura Xper F010/10 722011 Allura Xper F020 722012 Allura Xper FD20 biplane 722013 Allura Xper FDlO OR Table 722014 Allura Xper FD20 OR Table 722015 Allura Xper FDl0/10 OR Table 722019 Allura Xper FD20 Biplane OR Table 722020 Allura Xper FD10 OR Table 722022 Allura Xper FD20 OR Table 722023 Allura Xper FD10/10 OR Table 722024 Altura Xper FD20 Biplane OR Table 722025 Allura Xper FD10 722026 Allura Xper FD10/10 722027 Altura Xper FD20/10 biplane 722029 Allura Xper FD10 OR Table 722033 Allura Xper FD10/10 OR Table 722034 Allura Xper FD20/10 OR Table 722036 Allure Xper F020/20 722038 Altura Xper FD20/20 biplane OR Table 722039
Recall
- Recall Number
- Z-0724-2022
- Event Number
- 89375
- Firm
- Philips North America Llc
- FEI Number
- 3016618143
- Product Code
- OWB
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- November 24, 2021
- Address
- 222 Jacobs St, Cambridge, MA, 02141-2289
Description
Philips Allura Xper, UNIQ(Allura Xper system-China) Interventional Fluoroscopic X-ray System products with a Laird chiller WITHOUT an extended drip tray: Model Number: 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722029, 722033, 722034, 722036, 722038, 722039. Allura Xper F010 {C+F). 722003 Allura Xper F010/10 722005 Allura Xper F020 722006 Allura Xper F020/10 and F020/20 722008 Allura Xper F010 722010 Altura Xper F010/10 722011 Allura Xper F020 722012 Allura Xper FD20 biplane 722013 Allura Xper FDlO OR Table 722014 Allura Xper FD20 OR Table 722015 Allura Xper FDl0/10 OR Table 722019 Allura Xper FD20 Biplane OR Table 722020 Allura Xper FD10 OR Table 722022 Allura Xper FD20 OR Table 722023 Allura Xper FD10/10 OR Table 722024 Altura Xper FD20 Biplane OR Table 722025 Allura Xper FD10 722026 Allura Xper FD10/10 722027 Altura Xper FD20/10 biplane 722029 Allura Xper FD10 OR Table 722033 Allura Xper FD10/10 OR Table 722034 Allura Xper FD20/10 OR Table 722036 Allure Xper F020/20 722038 Altura Xper FD20/20 biplane OR Table 722039
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Philips Issued Medical Device Correction letter dated 11/24/21 to U.S.A, Customers via certified mailing from the United States Postal Service (USPS). Letters outside of the U.S.A. will be distributed through the Philips Markets Organizations. Letter states reason for recall, health risk and action to take: Philips will inspect the affected systems for leaks and repair as required. In addition, Philips will install an extended drip tray to prevent damage from any future leaks. This service will be provided free of charge for affected systems. A Philips service representative will contact customers with affected devices to arrange for the service. If you need any further information or support concerning this issue, please contact your local Philips representative: 1-800-722-9377 (reference FCC 72200484).
US Nationwide Foreign: Argentina Austria Bangladesh Brazil Chile China Denmark Dominican Republic France Germany Greece India Indonesia Iran Israel Italy Japan Korea, Republic of Malaysia Mexico Netherlands Philippines Russian Federation Saudi Arabia South Africa Spain Sri Lanka Switzerland Taiwan Thailand Turkey United Kingdom Virgin Islands (U.S.)
USA: 63 systems R.O.W.: 127 systems