FDA Recall Terminated

Endopath 5 mm Graspers with Ratchet Handles (product code 5DSG).

Recall: Z-0724-05 · Initiated December 15, 2004

Recall

Recall Number
Z-0724-05
Event Number
30727
Firm
Ethicon Endo-Surgery
FEI Number
1527736
Product Code
MDM
Status
Terminated
Root Cause
Other
Initiated
December 15, 2004
Posted
April 27, 2005
Terminated
May 30, 2012
Address
4545 Creek Rd, Cincinnati, OH, 45242-2803

Description

Endopath 5 mm Graspers with Ratchet Handles (product code 5DSG).

Reason

Due to tolerance issues with the device, pins can work free from their respective holes. The result during use could be separation of the handles that could lead to compromised ratchet button functionality. Loss of ratchet button functionality may lead to the end effector being locked in the open or closed position.

Action

The recalling firm sent letters to hospitals and distributors, dated 12/14/04.

Distribution

The devices were distributed to hospitals and distributors located throughout the United States and to foreign consignees located in Saudi Arabia, Austria, Egypt, United Kingdom, Germany, France, Lebanon, The Netherlands, Spain, Pakistan, South Africa, Italy, United Arab Emirate, Czech Republic, Sweden, Switzerland, Greece, India, Portugal, Slovenia, Hungary, Palestine, and Kuwait.

Quantity

29,141 devices.