MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429
Recall
- Recall Number
- Z-0722-2024
- Event Number
- 93673
- Firm
- Cook Medical Llc
- FEI Number
- 3007965706
- Product Code
- MQG
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- December 7, 2023
- Posted
- January 12, 2024
- Address
- 1025 W Acuff Rd, Bloomington, IN, 47404-9295
Description
MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429
The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.
The firm notified affected customers on December 7, 2023 via URGENT Medical Device Correction letters. Per the letter, customers can continue to use their MINC+ device whilst awaiting the defect correction, but Cook advises customers to be vigilant and monitor for device alerts. If the device produces audible and visual alerts to indicate an error with temperature, customers must immediately move any dishes to another incubator. If no other incubators are available, the device can be reset to normal operation by switching off mains power to the device for ten seconds then turning the power back on. The software/firmware update will undergo verification testing including compliance testing prior to release. The updated software/firmware is expected to be available on 01 February 2024. An authorized service agent will contact customers to arrange for the device to be corrected at their facility. The service agent will be updating the software/firmware of the device.
Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM
14 US; 101 OUS