FDA Recall Terminated

Stryker CBCII 1/8in. Double Trocar with quick disconnect , sterile, Stryker Instuments, Kalamazoo, MI, REF225-28-85S1. Blood recovery system used post-operatively to collect, filter and allow for reinfusion of autologous blood.

Recall: Z-0722-2010 · Initiated November 12, 2009

Recall

Recall Number
Z-0722-2010
Event Number
54149
Firm
Stryker Instruments Div. of Stryker Corporation
FEI Number
1811755
Product Code
cac
Status
Terminated
Root Cause
Packaging process control
Initiated
November 12, 2009
Posted
January 29, 2010
Terminated
October 25, 2010
Address
4100 E. Milham Ave., Kalamazoo, MI, 49001

Description

Stryker CBCII 1/8in. Double Trocar with quick disconnect , sterile, Stryker Instuments, Kalamazoo, MI, REF225-28-85S1. Blood recovery system used post-operatively to collect, filter and allow for reinfusion of autologous blood.

Reason

The product may not be sterile due to package not being intact.

Action

Stryker issued a "Medical Device Recall Notification" via email to foreign consignees dated November 12, 2009. Stryker further issued recall letters via overnight mail to domestic consginees beginning November 20, 2009. These notifications informed consignees of the affected devices instructing them to destroy all affected product and return a completed Business Reply Form to the firm. For further information, contact Stryker Instruments at 1-800-800-4236 extension 3808.

Distribution

Worldwide Distribution.

Quantity

79,617