Bard CritiCore Monitor, Software Version 2.1, Catalog number 000002N, C.R. Bard, Inc., Covington, GA 30014. Intended for multiple uses as it is an electronic battery-operated monitor used in conjunction with a single-use disposable urine collection container and/or a temperature sensing Foley catheter

Recall: Z-0719-2011 · Initiated November 17, 2010

Recall

Recall Number
Z-0719-2011
Event Number
57291
Firm
C.R. Bard, Inc., Urological Division
FEI Number
1018233
Product Code
EXS
Status
Terminated
Root Cause
Other
Initiated
November 17, 2010
Posted
December 21, 2010
Terminated
February 2, 2012
Address
8195 Industrial Blvd Ne, Covington, GA, 30014-1497

Description

Bard CritiCore Monitor, Software Version 2.1, Catalog number 000002N, C.R. Bard, Inc., Covington, GA 30014. Intended for multiple uses as it is an electronic battery-operated monitor used in conjunction with a single-use disposable urine collection container and/or a temperature sensing Foley catheter

Reason

Error can occur in the reading of the Present Interval Output (PIO) reading when the device is subjected to bumping or forceful replacement of the fluid container.

Action

Consignees were notified by an Urgent: Product Advisory Letter and Medica Device Recall letter on/about November 18, 2010. The letter identified the affected product, stated the reason for recall, the interim precautions, and actions to be taken. The letter requested acknowledgement of the recall and to return the entire affected product. Each consignee that is a distributor was instructed to forward the letter to their consignees to retrieve relevant product from those locations. Affected product received by Bard Medical will be reworked with the software upgrade and place back into distribution. Consignees will either receive a software upgrade on site by a third party or replacement product with the software upgrade. For international distribution, the hospital will be informed of this issue through Bard affiliate and Notified Body Bard Europe and managed according to the requirement. If there are questions or assistance is needed, customers can contact BD at 1-770-784-6969.

Distribution

Worldwide Distribution -- U.S., Netherlands, and Japan.

Quantity

234 units