Deltec branded CADD-Prizm PCS II ambulatory infusion pump and the CADD-Prizm VIP system ambulatory infusion pumps; (Both) Model 6101; Order #21-8861-01 (PCS II) and 21-8821-01 (VIP). Smiths Medical MD, Inc., St. Paul, MN 55112 USA; RX ONLY.

Recall: Z-0718-2007 · Initiated January 19, 2007

Recall

Recall Number
Z-0718-2007
Event Number
37486
Firm
Smiths Medical MD, Inc.
FEI Number
2183502
Product Code
FRN
Status
Terminated
Root Cause
Other
Initiated
January 19, 2007
Posted
April 5, 2007
Terminated
May 25, 2007
Address
1265 Grey Fox Rd, Saint Paul, MN, 55112-6929

Description

Deltec branded CADD-Prizm PCS II ambulatory infusion pump and the CADD-Prizm VIP system ambulatory infusion pumps; (Both) Model 6101; Order #21-8861-01 (PCS II) and 21-8821-01 (VIP). Smiths Medical MD, Inc., St. Paul, MN 55112 USA; RX ONLY.

Reason

The year of manufacture had been incorrectly entered as 1980 into the pumps software during the manufacture process. As a result the pump may go into an immediate 'Clock battery is Low/Service Immediately' or 'Clock Battery needs service soon' alarm message indicating that the pump requires servicing.

Action

All consignees who have received the affected CADD-Prizm PCS II ambulatory infusion pumps and the CADD-Prizm VIP System ambulatory infusion pumps were contacted via telephone by February 1, 2007. The problem will be explained to the consignee and they will be asked to return the pumps to Smiths Medical MD in order for the problem to be corrected.

Distribution

Nationwide.

Quantity

27