TransportAir, Accessory to an Air Bed, Auxiliary Blower Unit, manufactured by Kinetic Concepts, Inc.
Recall: Z-0718-03
·
Initiated June 1, 1997
Recall
- Recall Number
- Z-0718-03
- Event Number
- 25201
- Firm
- Kinetic Concepts, Inc
- FEI Number
- 1625774
- Product Code
- INX
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 1, 1997
- Posted
- April 10, 2003
- Terminated
- July 8, 2003
- Address
- 4958 Stout Drive, San Antonio, TX, 78219
Description
TransportAir, Accessory to an Air Bed, Auxiliary Blower Unit, manufactured by Kinetic Concepts, Inc.
Reason
Battery out-gassing due to battery over charging and/or secondary failure mode of deep-discharge.
Action
In approximately November 1997 the firm implemented the TransportAir retrograde for all active units.
Distribution
Nationwide
Quantity
4264