Arrow Berman Angiographic Catheter 60cm Item Number: AI-07130-J
Recall
- Recall Number
- Z-0716-2020
- Event Number
- 84399
- Firm
- Arrow International Inc
- FEI Number
- 3015859709
- Product Code
- DYG
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- November 15, 2019
- Terminated
- June 19, 2020
- Address
- 2400 Bernville Rd, Reading, PA, 19605-9607
Description
Arrow Berman Angiographic Catheter 60cm Item Number: AI-07130-J
Potential leak in the balloon extension line. If a leak was present and went undetected during pre-use testing as outlined in the IFU, this would cause the balloon to deflate slowly following placement. If the leak went undetected by attending clinicians during use, it could result in contrast injection into the vascular or cardiac tissue which could result in serious injury or death
Arrow International issued letters on 11/15/19 stating reason for recall, health risk, and action to take: immediately discontinue use and quarantine any products with the product codes and lot numbers listed above. 2. To acknowledge receipt of this Urgent Medical Device Recall Notice or to return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1- 855-419-8507, Attn: Customer Service or ema il [email protected]. your local sales representative or Customer Service at 1-866-396-2111. .
U.S. Nationwide, Canada
1 unit