FDA Recall Terminated

Allura CV20; Model Number: 722031 The Allura CV2O is intended for physicians (e.g. cardiologists and radiologists), assisted by trained hospital staff (e.g. nurses and lab technicians), who are qualified to perform medical procedures on humans (having a maximum weight of 250 kg.) with probable internal diseases or injuries for: " Dedicated vascular and carotid imaging applications, including diagnostic and interventional procedures. " Cardiac imaging applications including diagnostics, interventional rocedures, pacemaker implantations and electrophysiology (EP). " Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.

Recall: Z-0715-2016 · Initiated November 3, 2015

Recall

Recall Number
Z-0715-2016
Event Number
72599
Firm
Philips Electronics North America Corporation
FEI Number
1218950
Product Code
OWB
Status
Terminated
Root Cause
Device Design
Initiated
November 3, 2015
Posted
January 30, 2016
Terminated
September 15, 2020
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

Allura CV20; Model Number: 722031 The Allura CV2O is intended for physicians (e.g. cardiologists and radiologists), assisted by trained hospital staff (e.g. nurses and lab technicians), who are qualified to perform medical procedures on humans (having a maximum weight of 250 kg.) with probable internal diseases or injuries for: " Dedicated vascular and carotid imaging applications, including diagnostic and interventional procedures. " Cardiac imaging applications including diagnostics, interventional rocedures, pacemaker implantations and electrophysiology (EP). " Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.

Reason

The use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.

Action

The firm, Philips, issued "URGENT-Field Safety Notices" dated 10/26/2015 and 11/19/2015 to its customers. The notices described the product, problem and actions to be taken. The customers were instructed to ensure the audible signal (X-ray buzzer) is always enabled, to activate only ONE fluoroscopy pedal at time, and to configure the footswitch to the default Philips configuration. Customers were also instructed to review the information with all members of your staff who need to be aware; retain a copy with the equipment IFU and complete and return the Confirmation Respond form to Mail Stop 4302, 3000 Minuteman Road, Andover, MA 01810 or email: [email protected]. Philips is tracking the delivery of these notices via certified return receipt. If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377.

Distribution

Worldwide Distribution: US (Nationwide) including Puerto Rico and countries of: Albania, Algeria, US Virgin Islands, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Democratic Congo, Costa Rica, Croatia, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.

Quantity

13297 in total