Medisystems 17 Gauge x 1" inch AV Fistula Needle Product Code: D9-2007MGLB Product Usage: Medisystems AVF with MasterGuard anti-stick needle protector is indicated for use as a vascular access device in hemodialysis or other extracorporeal procedures. It is intended for use in conjunction with a blood tubing set equipped with a standard medical male luer-locking connector
Recall
- Recall Number
- Z-0715-2013
- Event Number
- 63959
- Firm
- Medisystems a NX Stage Company
- FEI Number
- 3003464075
- Product Code
- FIE
- Status
- Terminated
- Root Cause
- Labeling mix-ups
- Initiated
- December 20, 2012
- Posted
- January 24, 2013
- Terminated
- October 10, 2013
- Address
- 350 Merrimack St, Lawrence, MA, 01843-1748
Description
Medisystems 17 Gauge x 1" inch AV Fistula Needle Product Code: D9-2007MGLB Product Usage: Medisystems AVF with MasterGuard anti-stick needle protector is indicated for use as a vascular access device in hemodialysis or other extracorporeal procedures. It is intended for use in conjunction with a blood tubing set equipped with a standard medical male luer-locking connector
Single needle packs in each case ( 10 needles per case) are mislabeled on the pouch as 15 Gauge needles instead of 17 Gauge
Medisystems sent a notification letter dated December 20, 2012 to affected customers. The letter identified the affected products, problem and actions to be taken. Customers were instructed to immediately open and inspect inventory for the affected products. Results of the inspection are to be provided to Medisystems and the product is to be returned to Medisystems. For questions call (978) 332-5973.
US Nationwide Distribution in the states of NY, PA and TN.
16 cases (15 cs x 250 ea) 4,000 needles