FDA Recall
Terminated
Draeger Infinity Acute Care System Monitoring Solution
Recall: Z-0714-2014
·
Initiated December 10, 2013
Recall
- Recall Number
- Z-0714-2014
- Event Number
- 67056
- Firm
- Draeger Medical Systems, Inc.
- FEI Number
- 2510954
- Product Code
- MHX
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- December 10, 2013
- Posted
- January 10, 2014
- Terminated
- May 20, 2015
- Address
- 3135 Quarry Rd, Telford, PA, 18969-1042
Description
Draeger Infinity Acute Care System Monitoring Solution
Reason
After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.
Action
An Urgent Medical Device Recall Letter, dated December 2013, was sent to Users. Once available Drager will update components of the IACS Monitoring Solution System.
Distribution
US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.
Quantity
439