FDA Recall Terminated

Draeger Infinity Acute Care System Monitoring Solution

Recall: Z-0714-2014 · Initiated December 10, 2013

Recall

Recall Number
Z-0714-2014
Event Number
67056
Firm
Draeger Medical Systems, Inc.
FEI Number
2510954
Product Code
MHX
Status
Terminated
Root Cause
Software design
Initiated
December 10, 2013
Posted
January 10, 2014
Terminated
May 20, 2015
Address
3135 Quarry Rd, Telford, PA, 18969-1042

Description

Draeger Infinity Acute Care System Monitoring Solution

Reason

After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.

Action

An Urgent Medical Device Recall Letter, dated December 2013, was sent to Users. Once available Drager will update components of the IACS Monitoring Solution System.

Distribution

US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.

Quantity

439