FDA Recall Terminated

Coloplast Elefant Suction/Irrigation Cannula for Laparoscopy with Tube, Part Numbers: ASP 165, ASP 170, Coloplast Corp., Minneapolis, MN 55411

Recall: Z-0714-2008 · Initiated November 12, 2007

Recall

Recall Number
Z-0714-2008
Event Number
45907
Firm
Coloplast Corp
FEI Number
2125050
Product Code
FQH
Status
Terminated
Root Cause
Process design
Initiated
November 12, 2007
Posted
April 3, 2008
Terminated
November 4, 2008
Address
1525 W River Rd, Minneapolis, MN, 55411-3430

Description

Coloplast Elefant Suction/Irrigation Cannula for Laparoscopy with Tube, Part Numbers: ASP 165, ASP 170, Coloplast Corp., Minneapolis, MN 55411

Reason

Lack of sterilization assurance: Sterilization of Elefant Suction-Irrigation Devices is not performed according to ISO 11135 standard.

Action

Coloplast sent customers an "Urgent: Recall Notification" letter on 11/12/07. The letter described that the sterilization of the products was not performed according to ISO 11135 standard. The notification requested the customers to return the product immediately and to send a fax or contact the company via email in regard to inventory at hand. A phone script for distributors also included an advice to discontinue dispensing the product and to fax or email in regard to inventory at hand.

Distribution

Nationwide, including Puerto Rico.

Quantity

559