FDA Recall Terminated

DePuy Spine VIPER POLY SCREW 5.5 10X100MM Tl Product Code: 1867-15-199 Spinal Implant Component.

Recall: Z-0713-2013 · Initiated November 7, 2012

Recall

Recall Number
Z-0713-2013
Event Number
63665
Firm
DePuy Spine, Inc.
FEI Number
1526439
Product Code
NKB
Status
Terminated
Root Cause
Component change control
Initiated
November 7, 2012
Posted
January 24, 2013
Terminated
December 24, 2013
Address
325 Paramount Dr, Raynham, MA, 02767-5199

Description

DePuy Spine VIPER POLY SCREW 5.5 10X100MM Tl Product Code: 1867-15-199 Spinal Implant Component.

Reason

Polyaxial screws 8x 80 and larger may disassociate when inserting in the ilium. VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head.

Action

DePuy Synthes Spine sent an Urgent Voluntary Product Recall Notification letter dated November 7, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory, quarantine the material and return remaining inventory. Customers were advised to complete the enclosed Business Reply Form and fax back to 508-828-3762 or email to [email protected]. For questions call Linda Regan @508-977-6606.

Distribution

Worldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI, IL, MD,MI, MO,NC, TN, TX and the country of Australia.

Quantity

10 units