FDA Recall Terminated

ECAT Emerge PET Scanners (distributed by Siemens Medical Solutions USA, Inc.), ECAT ART PET Scanners (distributed by Siemens Medical Solutions USA, Inc.) and Sceptre PET Scanners (distributed by Hitachi Medical Systems of America)

Recall: Z-0712-06 · Initiated March 28, 2006

Recall

Recall Number
Z-0712-06
Event Number
35001
Firm
Siemens Medical Solutions USA Inc
FEI Number
1034973
Product Code
KPS
Status
Terminated
Root Cause
Other
Initiated
March 28, 2006
Posted
April 4, 2006
Terminated
December 22, 2009
Address
810 Innovation Dr, Knoxville, TN, 37932-2562

Description

ECAT Emerge PET Scanners (distributed by Siemens Medical Solutions USA, Inc.), ECAT ART PET Scanners (distributed by Siemens Medical Solutions USA, Inc.) and Sceptre PET Scanners (distributed by Hitachi Medical Systems of America)

Reason

There is a potential for inaccurate, elevated SUVs (standardized uptake values). On follow-up scans, the elevated SUV could be misinterpreted, and inappropriately impact a patient's treatment plan.

Action

Written notification was sent out beginning 03/28/2006 to end users to advise them of the problem. The notification also included specific work instructions to aid the end user until the patch is available and the upgrade could be performed by a service representative.

Distribution

KS, WA, CA, OH, FL, TX, Germany, Great Britain, Canada Switzerland, Turkey, Argentina, Austria, France, & Spain

Quantity

57 devices