DCTM Digital Ceiling Tube Mount, model DCTM; a fully counter-balanced radiographic x-ray tube suspension system designed to facilitate a wide range of procedures; Manufactured by Del Medical Imaging Corporation, Franklin Park, IL 60131 U.S.A.
Recall
- Recall Number
- Z-0712-04
- Event Number
- 28372
- Firm
- Del Medical Systems Group
- FEI Number
- 1418964
- Product Code
- IYB
- Status
- Terminated
- Root Cause
- Other
- Initiated
- February 13, 2004
- Posted
- July 20, 2004
- Terminated
- September 30, 2004
- Address
- 11550 West King Street, Franklin Park, IL, 60131-1330
Description
DCTM Digital Ceiling Tube Mount, model DCTM; a fully counter-balanced radiographic x-ray tube suspension system designed to facilitate a wide range of procedures; Manufactured by Del Medical Imaging Corporation, Franklin Park, IL 60131 U.S.A.
There is a potentially defective worm gear in the tension adjusting mechanism of the balancer used in the Digital Ceiling Mounted Tube Support System for overhead x-ray units.
Del Medical sent a Field Correction Recall letter dated 2/13/04 to all dealers/assemblers receiving DCTM''s during the affected time frame, alerting them to potential defective worm gear. The accounts were asked to locate the systems listed in the attachment to the letter, and alert the end users to take the units out of service until they can be inspected. Follow-up letters were sent via registerd mail on 2/19/04, reiterating the 2/13/04 letter and providing the accounts with instructions for inspection of the balancers. If the worm gear fails the test, the dealers were instructed to contact Del Medical for replacement of the balancer and assure that the unit is not placed back in service until the balancer has been replaced.
Nevada, Ohio, Tennessee, Nebraska, Arizona, Maryland, Wisconsin, Florida, Texas, Kansas, Oklahoma, Illinois, Colorado, California, Minnesota, Michigan, Alabama, Rhode Island and internationally to Canada, Panama and Pakistan
57 units