DePuy Spine VIPER POLY SCREW 5.5 10 X 70MM Tl Product Code: 1867-15-170 Spinal Implant Component.
Recall
- Recall Number
- Z-0711-2013
- Event Number
- 63665
- Firm
- DePuy Spine, Inc.
- FEI Number
- 1526439
- Product Code
- NKB
- Status
- Terminated
- Root Cause
- Component change control
- Initiated
- November 7, 2012
- Posted
- January 24, 2013
- Terminated
- December 24, 2013
- Address
- 325 Paramount Dr, Raynham, MA, 02767-5199
Description
DePuy Spine VIPER POLY SCREW 5.5 10 X 70MM Tl Product Code: 1867-15-170 Spinal Implant Component.
Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head
DePuy Synthes Spine sent an Urgent Voluntary Product Recall Notification letter dated November 7, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory, quarantine the material and return remaining inventory. Customers were advised to complete the enclosed Business Reply Form and fax back to 508-828-3762 or email to [email protected]. For questions call Linda Regan @508-977-6606.
Worldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI, IL, MD,MI, MO,NC, TN, TX and the country of Australia.
2 units