LIFEPAK 20 defibrillator/monitor
Recall
- Recall Number
- Z-0711-06
- Event Number
- 34947
- Firm
- Medtronic Emergency Response Systems, Inc.
- FEI Number
- 3015876
- Product Code
- LDD
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- March 22, 2006
- Posted
- April 4, 2006
- Terminated
- September 6, 2011
- Address
- 11811 Willows Rd Ne, Redmond, WA, 98052-2003
Description
LIFEPAK 20 defibrillator/monitor
Therapy connectors may have connector sockets that do not meet design specs - short drill depth. Interruption of the connections may cause unexpected, intermittent or complete loss of pacing and defibrillation functions.
On 3/22/06, the firm began sending out letters dated March 2006, and flagged URGENT MEDICAL DEVICE CORRECTION. The letters advised customers of the issue and that a service representative would visit and inspect the therapy connector sockets, and replace the therapy connector and the mating cable if short drill sockets are found.
Nationwide, and Canada, Netherlands, Germany, Australia, Guam, Hong Kong, New Zealand, New Caledonia, Argentina, Brazil, Chile, Costa Rica, Mexico, Uruguay, Venezuela. Domestic distribution includes hospitals, clinics,
2678 devices