FDA Recall Open, Classified

mint Lesion, Software Versions: 3.10.0 and 3.10.1.

Recall: Z-0710-2025 · Initiated November 15, 2024

Recall

Recall Number
Z-0710-2025
Event Number
95880
Firm
Mint Medical GmbH Friedrich-Ebert-Str. 2 Dossenheim Germany
FEI Number
3011706018
Product Code
LLZ
Status
Open, Classified
Root Cause
Software design
Initiated
November 15, 2024
Posted
December 20, 2024

Description

mint Lesion, Software Versions: 3.10.0 and 3.10.1.

Reason

Some software versions have a malfunction where they may show incorrect orientation labels for a specific subset of DICOM images.

Action

On 11/13/2024 and 11/15/2024 "URGENT FIELD SAFETY NOTICE" letters were emailed to customers. US customers on 11/13/2024 and OUS customers on 11/15/2024. Actions to be taken by the user: Please read this information carefully and assess whether you are using an affected product version and if DICOM series with differing patient orientation may be used in your organization (Note: mint Lesion" shows a notice above the images when multi-frame images are displayed (see Figures 1 and 2). If that is the case, the malfunction may occur in your system. Please be aware that displayed orientation markers may indicate a wrong patient orientation. Verify the patient orientation based on anatomical landmarks that you can see in the images, when you rely on patient orientation for reporting findings (e.g., "anomaly in left kidney"). If possible, do not assess DICOM data that may lead to the malfunction with mint Lesion" . If you suspect that your organization uses DICOM files with the described parameters and that they have already been read using mint Lesion" , you can contact Mint Medical support ([email protected]) to request analysis of your mint Lesion" instance. If you believe that this failure could have occurred in past use of mint Lesion", please review the potentially affected radiological reports in your reporting application and take the necessary steps to correct them. Please inform Mint Medical if this may have led to any patient harm in your organization. Actions being taken by the manufacturer The error will be corrected with a software update. Mint Medical Support will contact you when the update is available to schedule the installation of the update on your system. If you have any questions, email: [email protected] or call (+49) 6221 32 18018.

Distribution

Worldwide distribution: US (nationwide): AR, CA, FL, GA, MI, MN, NY, OH, PA, TN, TX, WI; and OUS (International) to countries of: Austria, Belgium, France, Germany, Italy, Netherlands, Spain, Switzerland, United Kingdom.

Quantity

18 units (7 US, 11 OUS)