FDA Recall Terminated

Axiom Artis, Model Numbers: 5904433, 5904441, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555373, 7727717, 7728350 & 7728392

Recall: Z-0708-2018 · Initiated June 6, 2017

Recall

Recall Number
Z-0708-2018
Event Number
79097
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
OWB
Status
Terminated
Root Cause
Device Design
Initiated
June 6, 2017
Terminated
May 16, 2019
Address
40 Liberty Blvd, Malvern, PA, 19355-1418

Description

Axiom Artis, Model Numbers: 5904433, 5904441, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555373, 7727717, 7728350 & 7728392

Reason

Potential for liquid entry

Action

On June 6, 2017, Siemens Healthineers distributed URGENT MEDICAL DEVICE CORRECTION notices to their customers via USPS Certified Mail. Siemens conducted a corrective field action in June 2015 to protect the Artis table against liquid entry at critical locations. However, even with this corrective action, if there are larger amounts of liquid on the table surface, they can still enter the table. Siemens Medical Solutions USA, Inc. will contact customers to arrange a date to perform the corrective action. Please feel free to contact the service organization for an earlier appointment by calling 610-448-6461.

Distribution

U.S. Nationwide