Axiom Artis, Model Numbers: 5904433, 5904441, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555373, 7727717, 7728350 & 7728392
Recall
- Recall Number
- Z-0708-2018
- Event Number
- 79097
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- OWB
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- June 6, 2017
- Terminated
- May 16, 2019
- Address
- 40 Liberty Blvd, Malvern, PA, 19355-1418
Description
Axiom Artis, Model Numbers: 5904433, 5904441, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555373, 7727717, 7728350 & 7728392
Potential for liquid entry
On June 6, 2017, Siemens Healthineers distributed URGENT MEDICAL DEVICE CORRECTION notices to their customers via USPS Certified Mail. Siemens conducted a corrective field action in June 2015 to protect the Artis table against liquid entry at critical locations. However, even with this corrective action, if there are larger amounts of liquid on the table surface, they can still enter the table. Siemens Medical Solutions USA, Inc. will contact customers to arrange a date to perform the corrective action. Please feel free to contact the service organization for an earlier appointment by calling 610-448-6461.
U.S. Nationwide