FDA Recall
Completed
Elekta Unity
Recall: Z-0706-2019
·
Initiated January 4, 2019
Recall
- Recall Number
- Z-0706-2019
- Event Number
- 81886
- Firm
- Elekta Limited Linac House Fleming Way Crawley United Kingdom
- FEI Number
- 3002806900
- Product Code
- IYE
- Status
- Completed
- Root Cause
- Under Investigation by firm
- Initiated
- January 4, 2019
Description
Elekta Unity
Reason
In rare circumstances, due to reasons such as inadequate hospital facility maintenance, water leakage, and weakening of speaker fixings during the lifetime of the system, the speakers can become dislodged and be pulled into the bore.
Action
The firm notified their consignees about the issue on 01/04/2019 and will install additional fixation.
Distribution
TX and WI, Germany, Denmark, Netherlands, United Kingdom
Quantity
2 units