FDA Recall
Terminated
Version 7 Software for wash program on Bio Tek Automated Microplate Washer Model ELx50 which performs the washing steps for various programmed mictotiter plate assays.
Recall: Z-0703-05
·
Initiated May 20, 2004
Recall
- Recall Number
- Z-0703-05
- Event Number
- 30931
- Firm
- Diasorin Inc.
- FEI Number
- 2182595
- Product Code
- JTC
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 20, 2004
- Posted
- April 13, 2005
- Terminated
- July 22, 2006
- Address
- 1951 Northwestern Ave S, Stillwater, MN, 55082-7536
Description
Version 7 Software for wash program on Bio Tek Automated Microplate Washer Model ELx50 which performs the washing steps for various programmed mictotiter plate assays.
Reason
When software was updated for the Washer program of the Bio Tek Automated Microplate Washer Model ELx50, the software did not automatically clear old versions of the wash program when new ones were added.
Action
Customers were notified by phone and informed of the problem and recommended users to use only the newer wash program which is identified is a 'W' in front of the program.
Distribution
CA, FL, NH, PA, SD, TX, WA
Quantity
9