Version 12 software program written for the Bio-Tek ELx800 automated microtiter plate reader with use of PRO-Trac Tacrolimus ELISA kit
Recall: Z-0702-05
·
Initiated April 12, 2004
Recall
- Recall Number
- Z-0702-05
- Event Number
- 30929
- Firm
- Diasorin Inc.
- FEI Number
- 2182595
- Product Code
- MLM
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 12, 2004
- Posted
- April 13, 2005
- Terminated
- July 22, 2006
- Address
- 1951 Northwestern Ave S, Stillwater, MN, 55082-7536
Description
Version 12 software program written for the Bio-Tek ELx800 automated microtiter plate reader with use of PRO-Trac Tacrolimus ELISA kit
Reason
The software program written for use in the automated microtiter plate reader (Bio-Tek ELx800) for use with the Pro-Trac II Tacrolimus ELISA kit was found to be faulty. Assay software program did not include the Calibrator Zero OD specification (STDO>=1.500) as specified in the product insert.
Action
Customers were sent notification letters and effectiveness check form. recommended users manually veify results using product insert specifications.
Distribution
CA, Washington DC, FL, KY & Canada