FDA Recall Terminated

Version 12 software program written for the Bio-Tek ELx800 automated microtiter plate reader with use of PRO-Trac Tacrolimus ELISA kit

Recall: Z-0702-05 · Initiated April 12, 2004

Recall

Recall Number
Z-0702-05
Event Number
30929
Firm
Diasorin Inc.
FEI Number
2182595
Product Code
MLM
Status
Terminated
Root Cause
Other
Initiated
April 12, 2004
Posted
April 13, 2005
Terminated
July 22, 2006
Address
1951 Northwestern Ave S, Stillwater, MN, 55082-7536

Description

Version 12 software program written for the Bio-Tek ELx800 automated microtiter plate reader with use of PRO-Trac Tacrolimus ELISA kit

Reason

The software program written for use in the automated microtiter plate reader (Bio-Tek ELx800) for use with the Pro-Trac II Tacrolimus ELISA kit was found to be faulty. Assay software program did not include the Calibrator Zero OD specification (STDO>=1.500) as specified in the product insert.

Action

Customers were sent notification letters and effectiveness check form. recommended users manually veify results using product insert specifications.

Distribution

CA, Washington DC, FL, KY & Canada