FDA Recall Open, Classified

5 IN (13 cm) Appx 1.5 ml, Bag Spike Adapter, Spiros, REF: 20123-01; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/ Integrated Clave Drip Chamber, Clave, Spiros, REF: CH3035; 40" (102 cm) Appx 4.5 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Hanger, REF: CH3135; 14" (36 cm) Appx 4.8 ml, Bifuse Add-On Set w/Spiros w/Red Cap, Bag Spike, 3 Clamps (Blue, 2 White), Vented Cap, Drop-In Spiros w/Red Cap, REF: CH3156; 114" Bifuse Primary Set w/2 Spiros, Bag Spike, 2 20 Drop In-line Drip Chamber, Clave, Bag Hanger, REF: CH3240; 18 IN (46 cm) Appx 2.9 ml, Add-On Set w/Spiros, Clave Spike, Clamp, Cap, REF: CH3290; 27" (69 cm) Appx 5.2 ml, Bifuse Add-On Set w/15 Drop Drip Chamber, Spiros w/Red Cap, Check Valve, Vented Cap, Bag Hanger, REF: CH3294; Secondary IV Oncology Kit, REF: CH3322; 30 IN (76 cm) Appx 1.8 ml) Ext Set w/2 Spiros, 2 Red Caps, Clamp, REF: CH3327; 40" (102 cm) Appx 4.6 ml, Admin Set, 2 Spiros w/2 Red Caps, 20 Drop In-Line Drip Chamber, 2 Bag Hangers, REF: CH3373; Oncology Kit w/Spiros, Red Cap, Bag Spike w/Clave Additive Port, REF: CH3397; 45" (114 cm) Appx 5.8 ml, Ext Set w/MicroClave, 2 Clave, Spiros w/Red Cap, 2 Clamps, REF: CH3403; Oncology Kit w/20" (51 cm) Ext Set w/MicroClave, Graduated Connector, Transfer Set w/Clave Additive Port, Spiros, Red Cap, 13mm Vial Spike w/Clave, REF: CH3507; Oncology Kit w/Clave Universal Vented Vial Spike, Spinning Spiros w/Red Cap, REF: CH3546; 22" (56 cm) Add-On 150 ml Burette Set (MicroClave, No Shut-Off), w/Spiros, 2 Clamps, Vented Cap, REF: CH3554; 4.5 IN (11 cm) Appx 0.43 ml, Ext Set, Spiros w/Red Cap, REF: CH3559; Oncology Kit w/Spinning Spiros, C-Clip, Red Cap, REF: CH3581; 5 IN (13 cm) Appx 0.57 ml, Ext Set, Anti-Siphon Valve, Spiros, REF: CH3582; Oncology Kit w/Spinning Spiros w/Red Cap, C-Clip; Priming Cap, REF: CH3589; 77" (196 cm) Appx. 9.4 ml, Admin Set w/2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber, Bag Hanger, REF: CH3639; 22" (56 cm) Appx 5.6ml, Ext Set, Trifuse w/4 Clave Clear, Spiros w/Red Cap, 1.2 Micron Filter, 4 Clamps, REF: CH3770; 6.5 IN (17 cm) Appx 0.78 ml, Ext Set w/Clave Clear, Clave 4-Way Stopcock, Spiros, Red Cap, REF: CH3771; 47" (119 cm) 50 ml Diluent Set for Channel 2 w/20 Drop In-Line Drip Chamber, 2 Spiros, REF: CH4001; 47" (119 cm) 50 ml Diluent Set For Channel 2 w/Bag Spike w/MicroClave Additive Port, Spiros, REF: CH4009; 30" (76 cm) Appx 5.9 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, REF: Z7100; 30" (76 cm) 50 ml Diluent Set for Channel 2 w/Spiros, REF: CH4006; 60 IN(152cm)APPX 0.45ml,SMALLBORE EXT,MicroCLAVE,Spiros, REF: Z7040; 6" (15 cm) Appx 0.55 ml, Smallbore Trifuse Ext Set w/3 MicroClave, 3 Clamps, Spiros, REF: Z7151

Recall: Z-0701-2021 · Initiated August 31, 2020

Recall

Recall Number
Z-0701-2021
Event Number
86426
Firm
ICU Medical, Inc.
FEI Number
2025816
Product Code
FPA
Status
Open, Classified
Root Cause
Process control
Initiated
August 31, 2020
Address
951 Calle, Amanecer San Clemente, CA, 92673-6212

Description

5 IN (13 cm) Appx 1.5 ml, Bag Spike Adapter, Spiros, REF: 20123-01; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/ Integrated Clave Drip Chamber, Clave, Spiros, REF: CH3035; 40" (102 cm) Appx 4.5 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Hanger, REF: CH3135; 14" (36 cm) Appx 4.8 ml, Bifuse Add-On Set w/Spiros w/Red Cap, Bag Spike, 3 Clamps (Blue, 2 White), Vented Cap, Drop-In Spiros w/Red Cap, REF: CH3156; 114" Bifuse Primary Set w/2 Spiros, Bag Spike, 2 20 Drop In-line Drip Chamber, Clave, Bag Hanger, REF: CH3240; 18 IN (46 cm) Appx 2.9 ml, Add-On Set w/Spiros, Clave Spike, Clamp, Cap, REF: CH3290; 27" (69 cm) Appx 5.2 ml, Bifuse Add-On Set w/15 Drop Drip Chamber, Spiros w/Red Cap, Check Valve, Vented Cap, Bag Hanger, REF: CH3294; Secondary IV Oncology Kit, REF: CH3322; 30 IN (76 cm) Appx 1.8 ml) Ext Set w/2 Spiros, 2 Red Caps, Clamp, REF: CH3327; 40" (102 cm) Appx 4.6 ml, Admin Set, 2 Spiros w/2 Red Caps, 20 Drop In-Line Drip Chamber, 2 Bag Hangers, REF: CH3373; Oncology Kit w/Spiros, Red Cap, Bag Spike w/Clave Additive Port, REF: CH3397; 45" (114 cm) Appx 5.8 ml, Ext Set w/MicroClave, 2 Clave, Spiros w/Red Cap, 2 Clamps, REF: CH3403; Oncology Kit w/20" (51 cm) Ext Set w/MicroClave, Graduated Connector, Transfer Set w/Clave Additive Port, Spiros, Red Cap, 13mm Vial Spike w/Clave, REF: CH3507; Oncology Kit w/Clave Universal Vented Vial Spike, Spinning Spiros w/Red Cap, REF: CH3546; 22" (56 cm) Add-On 150 ml Burette Set (MicroClave, No Shut-Off), w/Spiros, 2 Clamps, Vented Cap, REF: CH3554; 4.5 IN (11 cm) Appx 0.43 ml, Ext Set, Spiros w/Red Cap, REF: CH3559; Oncology Kit w/Spinning Spiros, C-Clip, Red Cap, REF: CH3581; 5 IN (13 cm) Appx 0.57 ml, Ext Set, Anti-Siphon Valve, Spiros, REF: CH3582; Oncology Kit w/Spinning Spiros w/Red Cap, C-Clip; Priming Cap, REF: CH3589; 77" (196 cm) Appx. 9.4 ml, Admin Set w/2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber, Bag Hanger, REF: CH3639; 22" (56 cm) Appx 5.6ml, Ext Set, Trifuse w/4 Clave Clear, Spiros w/Red Cap, 1.2 Micron Filter, 4 Clamps, REF: CH3770; 6.5 IN (17 cm) Appx 0.78 ml, Ext Set w/Clave Clear, Clave 4-Way Stopcock, Spiros, Red Cap, REF: CH3771; 47" (119 cm) 50 ml Diluent Set for Channel 2 w/20 Drop In-Line Drip Chamber, 2 Spiros, REF: CH4001; 47" (119 cm) 50 ml Diluent Set For Channel 2 w/Bag Spike w/MicroClave Additive Port, Spiros, REF: CH4009; 30" (76 cm) Appx 5.9 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, REF: Z7100; 30" (76 cm) 50 ml Diluent Set for Channel 2 w/Spiros, REF: CH4006; 60 IN(152cm)APPX 0.45ml,SMALLBORE EXT,MicroCLAVE,Spiros, REF: Z7040; 6" (15 cm) Appx 0.55 ml, Smallbore Trifuse Ext Set w/3 MicroClave, 3 Clamps, Spiros, REF: Z7151

Reason

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Action

On 08/31/2020, Medical Device Recall notices were mailed to customers informing them to take the following actions: 1) Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) ICU Medical has some lots of unaffected product available today and is actively increasing the amount of available inventory. In the event specific product is unavailable, consider use of the non-spinning Spiros or the ChemoLock CSTD as alternatives. Please contact customer service for product availability. 4) Return affected product using the return label provided with this letter. Contact Stericycle at 1-888-871-7108 (MF, 8am-5pm ET) if you have not received a return label or require additional labels for returning the affected product. 5) If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product / Table 1 sections of this notification and ask them to contact Stericycle to obtain a response form. For further inquiries, please contact: Global Complaint Management, 1-844-654-7780 or [email protected]; Customer Service, 1-866-829-9025, option 8 or, [email protected], (M-F, 8am-6pm CT) On 10/07/2020, the Medical Device Recall notice, with an expanded affected device scope, was sent to customers.

Distribution

Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.

Quantity

55000