Natrelle Inspira Sillicone-Filled Breast Implants for the following Styles and Sizes: 7607 SSM, 7608 SSF, 7609 SSX, 7513 SRL, 7514 SRLP, 7515 SRM, 7516 SRF, 7517 SRX, 7626 SCL, 7627 SCLP, 7628 SCM, 7629 SCF, 7630 SCX: 7607 SSM/UDI: Natrelle Inspira SSM 140cc BI (US) 10888628032835, Natrelle Inspira SSM 175cc BI (US) 10888628032842, Natrelle Inspira SSM 195cc BI (US) 10888628032859, Natrelle Inspira SSM 210cc BI (US) 10888628032866, Natrelle Inspira SSM 240cc BI (US) 10888628032873, Natrelle Inspira SSM 255cc BI (US) 10888628032880, Natrelle Inspira SSM 275cc BI (US) 10888628032897, Natrelle Inspira SSM 295cc BI (US) 10888628032903, Natrelle Inspira SSM 310cc BI (US) 10888628032910, Natrelle Inspira SSM 330cc BI (US) 10888628032927, Natrelle Inspira SSM 345cc BI (US) 10888628032934, Natrelle Inspira SSM 360cc BI (US) 10888628032941, Natrelle Inspira SSM 375cc BI (US) 10888628032958, Natrelle Inspira SSM 405cc BI (US) 10888628032965, Natrelle Inspira SSM 445cc BI (US) 10888628032972, Natrelle Inspira SSM 485cc BI (US) 10888628032989, Natrelle Inspira SSM 520cc BI (US) 10888628032996, Natrelle Inspira SSM 560cc BI (US) 10888628033009, Natrelle Inspira SSM 600cc BI (US) 10888628033016, Natrelle Inspira SSM 640cc BI (US) 10888628033023, Natrelle Inspira SSM 685cc BI (US) 10888628033030, 7608 SSF/UDI: Natrelle Inspira SSF 180cc BI (US) 10888628033054, Natrelle Inspira SSF 200cc BI (US) 10888628033061, Natrelle Inspira SSF 220cc BI (US) 10888628033078, Natrelle Inspira SSF 240cc BI (US) 10888628033085, Natrelle Inspira SSF 265cc BI (US) 10888628033092, Natrelle Inspira SSF 295cc BI (US) 10888628033108, Natrelle Inspira SSF 325cc BI (US) 10888628033115, Natrelle Inspira SSF 335cc BI (US) 10888628033122, Natrelle Inspira SSF 345cc BI (US) 10888628033139, Natrelle Inspira SSF 365cc BI (US) 10888628033146, Natrelle Inspira SSF 385cc BI (US) 10888628033153, Natrelle Inspira SSF 415cc BI (US) 10888628033160, Natrelle Inspira SSF 450cc BI (US) 10888628033177, Natrelle Inspira SSF 485cc BI (US) 10888628033184, Natrelle Inspira SSF 520cc BI (US) 10888628033191, Natrelle Inspira SSF 560cc BI (US) 10888628033207, Natrelle Inspira SSF 605cc BI (US) 10888628033214, Natrelle Inspira SSF 650cc BI (US) 10888628033221, Natrelle Inspira SSF 695cc BI (US) 10888628033238, Natrelle Inspira SSF 745cc BI (US) 10888628033245, Natrelle Inspira SSF 770cc BI (US) 10888628033252, 7609 SSX/UDI: Natrelle Inspira SSX 200cc BI (US) 10888628033269, Natrelle Inspira SSX 225cc BI (US) 10888628033276, Natrelle Inspira SSX 255cc BI (US) 10888628033283, Natrelle Inspira SSX 285cc BI (US) 10888628033290, Natrelle Inspira SSX 310cc BI (US) 10888628033306, Natrelle Inspira SSX 340cc BI (US) 10888628033313, Natrelle Inspira SSX 375cc BI (US) 10888628033320, Natrelle Inspira SSX 400cc BI (US) 10888628033337, Natrelle Inspira SSX 420cc BI (US) 10888628033344, Natrelle Inspira SSX 445cc BI (US) 10888628033351, Natrelle Inspira SSX 470cc BI (US) 10888628033368, Natrelle Inspira SSX 495cc BI (US) 10888628033375, Natrelle Inspira SSX 525cc BI (US) 10888628033382, Natrelle Inspira SSX 545cc BI (US) 10888628033399, Natrelle Inspira SSX 560cc BI (US) 10888628033405, Natrelle Inspira SSX 580cc BI (US) 10888628033412, Natrelle Inspira SSX 615cc BI (US) 10888628033429, Natrelle Inspira SSX 650cc BI (US) 10888628033436, Natrelle Inspira SSX 700cc BI (US) 10888628033443, Natrelle Inspira SSX 750cc BI (US) 10888628033450, Natrelle Inspira SSX 800cc BI (US) 10888628033467, 7513 SRL/UDI: Natrelle Inspira SRL 110cc BI (US) 10888628006645, Natrelle Inspira SRL 125cc BI (US) 10888628006652, Natrelle Inspira SRL 140cc BI (US) 10888628006669, Natrelle Inspira SRL 170cc BI (US) 10888628006676, Natrelle Inspira SRL 200cc BI (US) 10888628006683, Natrelle Inspira SRL 230cc BI (US) 10888628006690, Natrelle Inspira SRL 260cc BI (US) 10888628006706, Natrelle Inspira SRL 290cc BI (US) 10888628006713, Natrelle Inspira SR
Recall
- Recall Number
- Z-0701-2020
- Event Number
- 84318
- Firm
- Allergan PLC
- FEI Number
- 2011068
- Product Code
- FTR
- Status
- Terminated
- Root Cause
- Error in labeling
- Initiated
- November 6, 2019
- Terminated
- July 9, 2021
- Address
- 2525 Dupont Dr, Irvine, CA, 92612-1531
Description
Natrelle Inspira Sillicone-Filled Breast Implants for the following Styles and Sizes: 7607 SSM, 7608 SSF, 7609 SSX, 7513 SRL, 7514 SRLP, 7515 SRM, 7516 SRF, 7517 SRX, 7626 SCL, 7627 SCLP, 7628 SCM, 7629 SCF, 7630 SCX: 7607 SSM/UDI: Natrelle Inspira SSM 140cc BI (US) 10888628032835, Natrelle Inspira SSM 175cc BI (US) 10888628032842, Natrelle Inspira SSM 195cc BI (US) 10888628032859, Natrelle Inspira SSM 210cc BI (US) 10888628032866, Natrelle Inspira SSM 240cc BI (US) 10888628032873, Natrelle Inspira SSM 255cc BI (US) 10888628032880, Natrelle Inspira SSM 275cc BI (US) 10888628032897, Natrelle Inspira SSM 295cc BI (US) 10888628032903, Natrelle Inspira SSM 310cc BI (US) 10888628032910, Natrelle Inspira SSM 330cc BI (US) 10888628032927, Natrelle Inspira SSM 345cc BI (US) 10888628032934, Natrelle Inspira SSM 360cc BI (US) 10888628032941, Natrelle Inspira SSM 375cc BI (US) 10888628032958, Natrelle Inspira SSM 405cc BI (US) 10888628032965, Natrelle Inspira SSM 445cc BI (US) 10888628032972, Natrelle Inspira SSM 485cc BI (US) 10888628032989, Natrelle Inspira SSM 520cc BI (US) 10888628032996, Natrelle Inspira SSM 560cc BI (US) 10888628033009, Natrelle Inspira SSM 600cc BI (US) 10888628033016, Natrelle Inspira SSM 640cc BI (US) 10888628033023, Natrelle Inspira SSM 685cc BI (US) 10888628033030, 7608 SSF/UDI: Natrelle Inspira SSF 180cc BI (US) 10888628033054, Natrelle Inspira SSF 200cc BI (US) 10888628033061, Natrelle Inspira SSF 220cc BI (US) 10888628033078, Natrelle Inspira SSF 240cc BI (US) 10888628033085, Natrelle Inspira SSF 265cc BI (US) 10888628033092, Natrelle Inspira SSF 295cc BI (US) 10888628033108, Natrelle Inspira SSF 325cc BI (US) 10888628033115, Natrelle Inspira SSF 335cc BI (US) 10888628033122, Natrelle Inspira SSF 345cc BI (US) 10888628033139, Natrelle Inspira SSF 365cc BI (US) 10888628033146, Natrelle Inspira SSF 385cc BI (US) 10888628033153, Natrelle Inspira SSF 415cc BI (US) 10888628033160, Natrelle Inspira SSF 450cc BI (US) 10888628033177, Natrelle Inspira SSF 485cc BI (US) 10888628033184, Natrelle Inspira SSF 520cc BI (US) 10888628033191, Natrelle Inspira SSF 560cc BI (US) 10888628033207, Natrelle Inspira SSF 605cc BI (US) 10888628033214, Natrelle Inspira SSF 650cc BI (US) 10888628033221, Natrelle Inspira SSF 695cc BI (US) 10888628033238, Natrelle Inspira SSF 745cc BI (US) 10888628033245, Natrelle Inspira SSF 770cc BI (US) 10888628033252, 7609 SSX/UDI: Natrelle Inspira SSX 200cc BI (US) 10888628033269, Natrelle Inspira SSX 225cc BI (US) 10888628033276, Natrelle Inspira SSX 255cc BI (US) 10888628033283, Natrelle Inspira SSX 285cc BI (US) 10888628033290, Natrelle Inspira SSX 310cc BI (US) 10888628033306, Natrelle Inspira SSX 340cc BI (US) 10888628033313, Natrelle Inspira SSX 375cc BI (US) 10888628033320, Natrelle Inspira SSX 400cc BI (US) 10888628033337, Natrelle Inspira SSX 420cc BI (US) 10888628033344, Natrelle Inspira SSX 445cc BI (US) 10888628033351, Natrelle Inspira SSX 470cc BI (US) 10888628033368, Natrelle Inspira SSX 495cc BI (US) 10888628033375, Natrelle Inspira SSX 525cc BI (US) 10888628033382, Natrelle Inspira SSX 545cc BI (US) 10888628033399, Natrelle Inspira SSX 560cc BI (US) 10888628033405, Natrelle Inspira SSX 580cc BI (US) 10888628033412, Natrelle Inspira SSX 615cc BI (US) 10888628033429, Natrelle Inspira SSX 650cc BI (US) 10888628033436, Natrelle Inspira SSX 700cc BI (US) 10888628033443, Natrelle Inspira SSX 750cc BI (US) 10888628033450, Natrelle Inspira SSX 800cc BI (US) 10888628033467, 7513 SRL/UDI: Natrelle Inspira SRL 110cc BI (US) 10888628006645, Natrelle Inspira SRL 125cc BI (US) 10888628006652, Natrelle Inspira SRL 140cc BI (US) 10888628006669, Natrelle Inspira SRL 170cc BI (US) 10888628006676, Natrelle Inspira SRL 200cc BI (US) 10888628006683, Natrelle Inspira SRL 230cc BI (US) 10888628006690, Natrelle Inspira SRL 260cc BI (US) 10888628006706, Natrelle Inspira SRL 290cc BI (US) 10888628006713, Natrelle Inspira SR
The firm has became aware that 2D barcodes on breast implants and sizers may be unreadable by GS-1 configured scanners. These 2D barcodes contain a group separators between fixed-length element string in the UDI data chain of the 2D barcode, which is not formatted to the GS1 standard.
Allergan sent out the field notification to consignees via email on 11/06/19. For consignees whom email was unavailable, a field notification letter was sent via overnight mail with tracking to all consignees.In addition to informing them about the correction, the notification requested consignees do the following: 1. You may continue to use these products. If your facility uses GS-1 configured scanners and the 2D barcode cannot be scanned, the UDI information can be entered manually by using the human readable information that appears next to the 2D barcode.: 1. To enter the UDI, For the example below, the complete UDI would be: (01) 10888628004221 (17) 20220512 (21) 23260389. 2. Verify the number was entered correctly. 3. If you received this as an email and did not provide a read receipt, please response directly to this email to confirm receipt. If you received this as a letter, please return the enclosed confirmation card. Your confirmation is an important part of our record keeping. 4. YOU DO NOT NEED TO RETURN ANY PRODUCT. 5. If you require further information or assistance, please contact Allergan Medical Information at 1-800-678-1605, Option 2 or [email protected]. To report product complaints, please contact Allergan Product Surveillance at 1-800-624-4261. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
US:AL, AR,AZ,CA,CO,DC,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,WA,WI,WY