FDA Recall Open, Classified

HUDSON RCI mBrace, ET Tube Holder with 4 Point Head Strap, REF DYNJMBRC4; tracheal tube fixation device

Recall: Z-0697-2025 · Initiated November 18, 2024

Recall

Recall Number
Z-0697-2025
Event Number
95821
Firm
MEDLINE INDUSTRIES, LP Northfield
FEI Number
1417592
Product Code
CBH
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
November 18, 2024
Posted
December 18, 2024
Address
3 Lakes Dr, Northfield, IL, 60093-2753

Description

HUDSON RCI mBrace, ET Tube Holder with 4 Point Head Strap, REF DYNJMBRC4; tracheal tube fixation device

Reason

Complaint defects have been noted regarding the adhesive elements of the device and the frame of the holder putting additional pressure on the patient's cheekbones and upper lip which may cause dermal injury depending on the device positioning.

Action

Medline Industries issued an URGENT MEDICAL DEVICE RECALL to its consignees of 11/18/2024 via USPS and email. The notice explained the issue, risk to user, and requested the destruction of the affected product. For questions, contact the Recall Department at 866-359-1704 or [email protected].

Distribution

US and Canada

Quantity

1464 units