FDA Recall Terminated

Fisher & Paykel Healthcare Adult Respiratory Ventilator Circuit-Dual Heated Catalogue Number RT210

Recall: Z-0696-06 · Initiated December 20, 2005

Recall

Recall Number
Z-0696-06
Event Number
34248
Firm
Fisher & Paykel Healthcare Inc
FEI Number
3000210616
Product Code
BTT
Status
Terminated
Root Cause
Other
Initiated
December 20, 2005
Posted
March 29, 2006
Terminated
April 19, 2012
Address
22982 Alcalde Drive, 101, Laguna Hills, CA, 92653

Description

Fisher & Paykel Healthcare Adult Respiratory Ventilator Circuit-Dual Heated Catalogue Number RT210

Reason

Risk of fire. The manufacturing process may have the potential to damage the insulation of the heater wire in circuits within the specified lot range. This may lead to a potential risk of fire.

Action

On December 20th, 2005 the firm sent the notification letter instructing consignees to check and destroy any stock in the affected lot range. A fax-back Device Recall Response Form was included with the letter and will be used for the company''s effectiveness checks.

Distribution

Nationwide in AR, CA, FL, IL, LA, MN, NC, OH, SD, TX, UT, & VA. International distribution

Quantity

2175 boxes (21,750 pcs)