Medtronic Soletra 7426, Quadripolar Multi-Programmable Neurostimulator for Deep Brain Stimulation, Rx only, Medtronic, Inc., Minneapolis, MN 55432-5604 USA (The Medtronic Activa System in an implantable, multiprogrammable quadripolar system that delivers electrical stimulation to selected areas of the brain)
Recall
- Recall Number
- Z-0694-2008
- Event Number
- 45847
- Firm
- Medtronic Neuromodulation
- FEI Number
- 2182207
- Product Code
- MHY
- Status
- Terminated
- Root Cause
- Packaging change control
- Initiated
- October 1, 2007
- Posted
- February 5, 2008
- Terminated
- May 3, 2012
- Address
- 800 53rd Ave NE PO Box 1250, Minneapolis, MN, 55440-1250
Description
Medtronic Soletra 7426, Quadripolar Multi-Programmable Neurostimulator for Deep Brain Stimulation, Rx only, Medtronic, Inc., Minneapolis, MN 55432-5604 USA (The Medtronic Activa System in an implantable, multiprogrammable quadripolar system that delivers electrical stimulation to selected areas of the brain)
Cessation of Therapy: separation of internal connections (lifted wirebonds) between the electronic circuit and battery may lead to sudden cessation of therapy.
A letter "Medtronic Urgent: Device Recall" dated October 2007 was sent to Healthcare Professionals. The letter described the device defect, product involved, root cause, probability of occurrence, recommendations and also provided information for assistance. At the time of the recall all of the product remaining in date for use (not expired) had been implanted, therefore Medtronic provided Important Patient Management Information for the implanting physicians to use in patient care and decisions regarding patient explants.
Worldwide: USA including the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, WA, WI, WY, UT, VA, VT, and WV and countries of Australia , Austria , Belgium , Canada , Cyprus , Czech Republic , Denmark , Finland , France , Germany , Greece , Hong Kong , Hungary , Iceland , India , Iran , Israel , Italy , Japan , Latin America, Lebanon , The Netherlands , Norway , Poland , Portugal , Russian Federation , Saudi Arabia , South Africa , Spain , Sweden , Switzerland , Turkey , and The United Kingdom
1,652