FDA Recall
Terminated
Fisher & Paykel Healthcare Adult Respiratory Ventilator Circuit-Dual Heated Catalogue Number RT110
Recall: Z-0694-06
·
Initiated December 20, 2005
Recall
- Recall Number
- Z-0694-06
- Event Number
- 34248
- Firm
- Fisher & Paykel Healthcare Inc
- FEI Number
- 3000210616
- Product Code
- BTT
- Status
- Terminated
- Root Cause
- Other
- Initiated
- December 20, 2005
- Posted
- March 29, 2006
- Terminated
- April 19, 2012
- Address
- 22982 Alcalde Drive, 101, Laguna Hills, CA, 92653
Description
Fisher & Paykel Healthcare Adult Respiratory Ventilator Circuit-Dual Heated Catalogue Number RT110
Reason
Risk of fire. The manufacturing process may have the potential to damage the insulation of the heater wire in circuits within the specified lot range. This may lead to a potential risk of fire.
Action
On December 20th, 2005 the firm sent the notification letter instructing consignees to check and destroy any stock in the affected lot range. A fax-back Device Recall Response Form was included with the letter and will be used for the company''s effectiveness checks.
Distribution
Nationwide in AR, CA, FL, IL, LA, MN, NC, OH, SD, TX, UT, & VA. International distribution
Quantity
2699 boxes (26,990 pcs)