Galaxy System (GAL-001)
Recall
- Recall Number
- Z-0693-2025
- Event Number
- 95712
- Firm
- Noah Medical
- FEI Number
- 3021325287
- Product Code
- EOQ
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- October 25, 2024
- Posted
- December 18, 2024
- Address
- 1501 Industrial Rd, San Carlos, CA, 94070-4111
Description
Galaxy System (GAL-001)
Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.
On 10/25/24, the firm started hand delivering correction notices to customers informing them of the following: If your device is affected the firm will contact you to inspect and if necessary to replace the power supply. Firm contact: [email protected], 1-888-325-6624
US: CA, MA, PA, NJ, ID, CT, MO, NC, MD, IL, TN, VA, OH, IN, WV, NY