FDA Recall Open, Classified

Galaxy System (GAL-001)

Recall: Z-0693-2025 · Initiated October 25, 2024

Recall

Recall Number
Z-0693-2025
Event Number
95712
Firm
Noah Medical
FEI Number
3021325287
Product Code
EOQ
Status
Open, Classified
Root Cause
Device Design
Initiated
October 25, 2024
Posted
December 18, 2024
Address
1501 Industrial Rd, San Carlos, CA, 94070-4111

Description

Galaxy System (GAL-001)

Reason

Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.

Action

On 10/25/24, the firm started hand delivering correction notices to customers informing them of the following: If your device is affected the firm will contact you to inspect and if necessary to replace the power supply. Firm contact: [email protected], 1-888-325-6624

Distribution

US: CA, MA, PA, NJ, ID, CT, MO, NC, MD, IL, TN, VA, OH, IN, WV, NY

Quantity

26