FDA Recall Terminated

Medtronic Affinity, 38um Arterial Filter, Catalog number REF 351

Recall: Z-0693-2020 · Initiated November 14, 2019

Recall

Recall Number
Z-0693-2020
Event Number
84410
Firm
Medtronic Perfusion Systems
FEI Number
1000116158
Product Code
DTM
Status
Terminated
Root Cause
Process control
Initiated
November 14, 2019
Terminated
October 5, 2020
Address
7611 Northland Dr N, Brooklyn Park, MN, 55428-1088

Description

Medtronic Affinity, 38um Arterial Filter, Catalog number REF 351

Reason

Medtronic has identified an out of-specification condition exhibiting excess plastic (flash) in the arterial filter directly above the outlet port.

Action

Beginning 14-Nov-2019, Medtronic initiated an immediate Urgent Medical Device Recall via verbal communication to one (1) customer that was shipped the affected product according to Medtronic records. The customer was asked to identify and quarantine any unused affected product. The immediate communication was followed by a written urgent medical device recall notification beginning on 26-Nov-2019 to communicate the issue to the one (1) customer that received affected product. The consignee was asked to confirm receipt of urgent medical device recall notification via a Customer Confirmation Certificate. Medtronic is seeking the return of the affected product.

Distribution

CA

Quantity

24 units