FDA Recall Terminated

Ingenuity CT (Computed Tomography) System, 510 (k) #K033326, Model #728326, Serial Numbers: 300005, 300010, 300011, 300012, 300013, 300014, 300015, 300016, 300017, 300018, 300019, 300020 and 300021. Product Usage: The Ingenuity CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of X-Ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

Recall: Z-0693-2012 · Initiated November 11, 2011

Recall

Recall Number
Z-0693-2012
Event Number
60741
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
JAK
Status
Terminated
Root Cause
Device Design
Initiated
November 11, 2011
Posted
January 11, 2012
Terminated
February 22, 2013
Address
595 Miner Road, Cleveland, OH, 44143-2131

Description

Ingenuity CT (Computed Tomography) System, 510 (k) #K033326, Model #728326, Serial Numbers: 300005, 300010, 300011, 300012, 300013, 300014, 300015, 300016, 300017, 300018, 300019, 300020 and 300021. Product Usage: The Ingenuity CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of X-Ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

Reason

Component defect. When the user is operating the control panel, he/she may inadvertently activate the wrong switch on the control panel which could, in turn, cause the patient couch to move in an unwanted direction.

Action

Philips sent an 'URGENT - Medical Device Correction Ingenuity CT' letter dated November 15, 2011. The letter describes the affected product, problem, hazard involved and actions to be taken by the customer and user. In this case, the customers are advised that extra care should be taken when using the 8-way Gantry Panel switch so as to avoid inadvertently activating an unintended switch resulting in unwanted (couch) movement. The letter notifies the customers that a Philips representative will be visiting each customer site and installing a new control panel through a Field Change Order. For further information or support concerning this issue contact your local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customer Care Solutions Center at (1-800-722-9377, option 5; enter site ID or follow prompts).

Distribution

Worldwide Distribution - USA (nationwide) in the states of: LA, NY, OH, WV and the foreign countries of: Australia, Germany, Korea, India, Malaysiaa and New Zealand.

Quantity

13 system units were distributed (5 in U.S. & 8 overseas)