FDA Recall Terminated

Berchtold Chromophare Dual Monitor Carrier Arm or Dual Flat Panel, a component of Berchtold Chromophare Ceiling Mounted Surgical Light System.

Recall: Z-0692-2018 · Initiated January 26, 2018

Recall

Recall Number
Z-0692-2018
Event Number
79116
Firm
Stryker Communications
FEI Number
3002975342
Product Code
FSY
Status
Terminated
Root Cause
Process control
Initiated
January 26, 2018
Terminated
March 19, 2019
Address
571 Silveron Blvd, Flower Mound, TX, 75028

Description

Berchtold Chromophare Dual Monitor Carrier Arm or Dual Flat Panel, a component of Berchtold Chromophare Ceiling Mounted Surgical Light System.

Reason

The mounts on the monitor may have an insufficient (missing or incomplete) weld.

Action

The recalling firm issued letters dated 1/26/2018 via FedEx on 1/26/2018 explaining the issue and the potential adverse consequences.

Distribution

Distribution was made to medical facilities nationwide, including government and military distribution. Foreign distribution was made to Canada, Belgium, China, United Kingdom, UAE, France, Germany, Spain, Japan, Ecuador, and Russia.

Quantity

850