FDA Recall Terminated

Varian High Energy Linear Accelerator, C-Series Clinac, Reference/FSCA Identifier: CP-06611; Models Numbers: 600C, 600CD, 6EX, DBX, 2100C, 2100CD, 2300CD, 21EX, 23EX, DMX, DHX versions 2.x through 6.x; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA. The Varian High Energy Linear Accelerator is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Recall: Z-0692-2012 · Initiated November 28, 2011

Recall

Recall Number
Z-0692-2012
Event Number
60659
Firm
Varian Medical Systems, Inc. Oncology Systems
FEI Number
2916710
Product Code
IYE
Status
Terminated
Root Cause
Device Design
Initiated
November 28, 2011
Posted
January 11, 2012
Terminated
June 2, 2015
Address
911 Hansen Way, Palo Alto, CA, 94304-1028

Description

Varian High Energy Linear Accelerator, C-Series Clinac, Reference/FSCA Identifier: CP-06611; Models Numbers: 600C, 600CD, 6EX, DBX, 2100C, 2100CD, 2300CD, 21EX, 23EX, DMX, DHX versions 2.x through 6.x; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA. The Varian High Energy Linear Accelerator is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Reason

An event has been reported to Varian which entails excessive connector resistance, which caused the actual jaw positions to differ from the intended jaw positions without warning the operator.

Action

Varian Medical Systems sent an Urgent Medical Device Correction - Urgent Field Safety Notice dated November 28, 2011 ***Expanded to include product #2 on December 16, 2011***, to all affected customers. ***12-12-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to establish a comprehensive quality assurance program before beginning treatments for the day, quality assurance tests must be performed. Additionally, any discrepancies in imaging field sizes vs planned field size should be brought to the attention of the medical physicist for investigation before treatment continues. Customers were asked to complete the attached Proof of Notification or Receipt Verification Card once they have read the document and return it to Varian Medical Systems. For question regarding this recall: USA and Canada 1-888-827-4265 Europe +41 41 749 8844

Distribution

Worldwide Distribution

Quantity

6158 units total